Table 5: Short overview on push button functions
Function
Push but-
tons
→
press simultaneously for 5 s: Activate setting mode
[S1] + [S2]
→
press in setting mode for 3 s: Set 4 mA
→
press in setting mode for 6 s: Set 0 mA
→
press in operation for 3 s: Switch on/off LED end position signalling.
→
touch in end position: Reduce current value by 0.02 mA
[S1]
→
press in setting mode for 3 s: Set 20 mA
→
press in operation for 3 s: Switch on/off LED end position signalling.
→
touch in end position: Increase current value by 0.02 mA
[S2]
9.7.1.
Measuring range: set
For measuring range setting, voltage must be applied at the position transmitter.
Information
●
Both measuring ranges 0/4 – 20 mA and 20 – 0/4 mA (inverse operation) can
be set.
During setting process, the measuring range (normal or inverse operation) is
assigned to the end positions by push button S1/S2 assignment.
●
Setting mode activating clears the setting in both end positions and sets the
output current to a value of 3.5 mA. After activation, both end values (0/4 mA
and 20 mA) need to be reset.
●
In case of inadvertent incorrect adjustment, the settings can always be reset
by renewed activation of the setting mode (simultaneous pressing of [S1] and
[S2]).
Activate setting mode
1.
Press both push buttons [S1] and [S2] and hold down for 5 seconds:
➥
By pulsing double flashes, the LED indicates that the setting mode is correctly
activated:
➥
For any other LED flash sequence (single/triple flashing):
➔
Refer to <Faults
during commissioning>.
Set measuring range
2.
Operate valve in one of the end positions (OPEN/CLOSED).
3.
Set desired output current (0/4 mA or 20 mA):
→
for 4 mA: Hold down push button [S1] for approx. 3 seconds,
until LED is slowly blinking
.
→
for 0 mA: Hold down push button [S1] for approx. 6 seconds,
until LED is blinking fast
.
→
for 20 mA: Hold down push button [S2] for approx. 3 seconds,
until LED is illuminated
.
4.
Operate valve into opposite end position.
➥
The value set in end position (0/4 mA or 20 mA) does not change during travel
in setting mode.
5.
Perform setting in the second end position following the same steps.
29
SQ 05.2 – SQ 14.2/SQR 05.2 – SQR 14.2
Commissioning