![Aulisa Guardian Angel Rx GA1000 Instructions For Use Manual Download Page 65](http://html1.mh-extra.com/html/aulisa/guardian-angel-rx-ga1000/guardian-angel-rx-ga1000_instructions-for-use-manual_3016355065.webp)
62
Aulisa GA1000 Digital Vital Sign Monitoring System
Blood Oxygen Saturation
Display Range (SpO )
2
Pulse Rate Display Range
Accuracy
Blood Oxygen Saturation
Pulse Rate
1% to 100%
30 to 290 bpm
70-100% ± 3 digits
± 3%
Alarms
SpO
2
Default SpO Limit
2
Upper Limit
Lower Limit
Alarm Limit Range
Upper Limit
Lower Limit
Adjustment Step
Step Value
Pulse Rate
Default Pulse Rate Limit
Upper Limit
Lower Limit
Alarm Limit Range
Upper Limit
Lower Limit
Adjustment Step
Step Value
Off
85%
85-100%
50-95%
1% SpO
2
150 bpm
50 bpm
75-275 bpm
30-110 bpm
1 bpm