23
Use the following procedure to verify that the alarm systems are
working properly.
NOTE:Alarm systems should be checked before use.
1. Set up Aulisa GA1000 system according to instructions above.
(See “Setting up Aulisa GA1000 system” section on page 18 for more
information.)
2. Ensure there is a wireless connection established between the
Oximeter Box and Display Unit. (See “Device Pairing” section on
page 22 more information.)
3. Detach the sensor cable from Oximeter Box.
4. Verify that an alarm message is displayed and that an alarm audio is
generated. (See “Troubleshooting” section on page 39 if an alarm
message and audio signal is not generated.)
5. Press on the "PAUSE AUDIO" button to temporarily silence for
2 minutes.
6. After alarm signal is regenerated, press on the "AUDIO OFF" button
to silence permanently the alarm signal.
Use the procedure below to monitor the readings (SpO , pulse rate,
2
pulse amplitude) to verify that the device is functioning properly.
1. Set up Aulisa GA1000 system according to instructions above.
(See “Setting up Aulisa GA1000 system” section on page 18 for more
information.)
2. Press firmly on the application site to make sure it securely attached
to the finger.
3. Verify the Bluetooth connection status icon on the Display Unit is blue
and the status indicator on the Oximeter Box is blinking green.
4. Verify that a SpO reading is displayed, that a pulse rate value
2
appears, and that a pulse amplitude reading is displayed.