Page  16

English

Standard EN 60118-4 of March 2007 

This standard defines 4 very precise parameters. 

The following procedure is recommended:

The peak magnetic field must reach 400mA/m (integration time 0.125ms).

Frequency responses should not vary more than 3dB between 100Hz and 5000Hz.

The signal-to-noise ratio must be 47 dB (A-weighting).

The magnetic field must be uniform throughout the listening area.

The type-approval procedure 

Recommendation

Thanks to the law of February 2005, establishments receiving the public (ERP) are gradually equip-
ping themselves with magnetic induction loops in order to facilitate listening for people with hearing 
difficulties. To be effective these systems must be carefully calibrated to verify that the installation 
has been made in compliance with standard EN60118-4. We recommend the delivery of a certifi-
cate of conformity to this standard, according to the model enclosed, signed by the installer.

To allow a good listening it is necessary to check all these conditions.
The respect of the magnetic field intensity guarantees that the sound intensity is sufficient, the 
responses frequency control ensure that the entire sound spectrum is perceived in the same way 
and the signal-to-noise ratio measures the impact of interference magnetic disturbances on intelligi-
bility. Despite the installation of adequate equipment, many factors can disturb these parameters: 
the presence of metallic structures causes the magnetic field to drop, transformers or current at 
50Hz in the proximity cause magnetic noises. The size and the form of the room can make it difficult 
to meet the recommended values at all points.

Procedure for receiving 

magnetic loops

Room managers must ensure that, when installing any magnetic induction loop, the installer delivers 
this certificate of conformity. Even if the room manager does not have the competence to judge the 
result of the measurements, this "installation report" should be able to be consulted at any time for a 
possible control by a team of auditors.

About the room, determine the points where the measurements will be taken ( center, extremities, 

points near metal structures or noise sources).
Measurements by the installer, after the adjustment of the installed equipment, of the various 

parameters set by the standard.
Delivery of this document signed by the installer to the room manager.

Summary of Contents for BM-SCAN

Page 1: ...USER GUIDE H10878 Version 1 03 2018 Magnetic field strenght meter BM SCAN...

Page 2: ...magnetic loops in the appendix You can then complete the certificate of compliance ON Press and hold button A until the display lights up Off Press and hold button A until the unit turns off When the...

Page 3: ...display will show the RMS value on the first line and the MAX value on the second line Press button B to reset the MAX measurement Location of the receiver coil A button M A and control mode button M...

Page 4: ...standard revised in 2004 notes that any background noise level below 47dBA will result in excellent signal noise but levels below 32dBA are acceptable and meet the requirements of the standard If the...

Page 5: ...average signal levels average is de 12dB or 100ma m for hearing instruments equipped with the T position The methods below are two processes that can be used to confirm that the system correctly repro...

Page 6: ...ck First it can be used to monitor the loop program and gives you a weighted output signal that can be heard with standard headphones To change the headphone volume drag the Use Menu button to Menu an...

Page 7: ...scroll through the different modes until you see Line In Flat Response on the screen Measuring range 62dB to 9dB 0dB 400mA m Peak factor 3 Resolution 0 1dB resolution for higher levels 32dB Measuremen...

Page 8: ...es that the sound intensity is sufficient the responses frequency control ensure that the entire sound spectrum is perceived in the same way and the signal to noise ratio measures the impact of interf...

Page 9: ...2 Customer Installer quipement Location Room Serial N Comments I declare that the installation complies with the IEC 60118 4 standard Installer s signature Date 7 Existence of other systems nearby Ye...

Page 10: ...modifications without prior notice That is why technical specifications and the products physical configuration might differ from the illustrations Make sure you get the latest news and updates about...

Reviews: