BM-SCAN - 

Magnetic field strenght meter

Page  14

English

Notes and requirements of Standard 60118-4

The standard notes that in frequencies from 100Hz to 5000Hz the loop should also reproduce all 
signals. At a minimum, systems should be tested at 100Hz, 1KHz and 5KHz.

Record RMS level readings for each frequency.
As in method I, if the level does not vary by more than ± 3dB, the installed system meets the 
IEC specifications of the standard. This method was requested by the  technicians so that they 
could perform the test with an instrument without continuously adjusting the frequency source. 
It also facilitates testing in several locations.

In the menu, the following items can be set: backlight level, headphone output type, headphone 
volume, display units (dB, mG, uT) and switch-off delay (5 to 30min or none).

The 2-purpose headphone output jack: First, it can be used to monitor the loop program and 
gives you a weighted output signal that can be heard with standard headphones. To change 
the headphone volume, drag the Use / Menu button to Menu and use the Mode button to 
switch to the headphone volume. By pressing the "select" button, you will be directed to the 
volume control screen where the upper button increases the volume and the lower button 
decreases the volume. Once set, simply return to operation to save the setting. 
 
A complete spectrum output can be sent from this same connector which could then feed a 
spectrum analyzer. 
This would show the signal level at different frequencies and help confirm proper operation 
and evaluate the frequency and level of any interference. 
 
To switch from Weighted to Flat: With the display on, switch to Use / Menu in "Menu", press 
the "A" mode button once to go to the Headphone Jack setup screen, then, by pressing the 
"B" selection button, you can choose a weighted or flat spectrum.

7 - Headphone output

8 - Menu settings

BP-RMS : 

-22.5 dB

MAX : 

1000 Hz

Summary of Contents for BM-SCAN

Page 1: ...USER GUIDE H10878 Version 1 03 2018 Magnetic field strenght meter BM SCAN...

Page 2: ...magnetic loops in the appendix You can then complete the certificate of compliance ON Press and hold button A until the display lights up Off Press and hold button A until the unit turns off When the...

Page 3: ...display will show the RMS value on the first line and the MAX value on the second line Press button B to reset the MAX measurement Location of the receiver coil A button M A and control mode button M...

Page 4: ...standard revised in 2004 notes that any background noise level below 47dBA will result in excellent signal noise but levels below 32dBA are acceptable and meet the requirements of the standard If the...

Page 5: ...average signal levels average is de 12dB or 100ma m for hearing instruments equipped with the T position The methods below are two processes that can be used to confirm that the system correctly repro...

Page 6: ...ck First it can be used to monitor the loop program and gives you a weighted output signal that can be heard with standard headphones To change the headphone volume drag the Use Menu button to Menu an...

Page 7: ...scroll through the different modes until you see Line In Flat Response on the screen Measuring range 62dB to 9dB 0dB 400mA m Peak factor 3 Resolution 0 1dB resolution for higher levels 32dB Measuremen...

Page 8: ...es that the sound intensity is sufficient the responses frequency control ensure that the entire sound spectrum is perceived in the same way and the signal to noise ratio measures the impact of interf...

Page 9: ...2 Customer Installer quipement Location Room Serial N Comments I declare that the installation complies with the IEC 60118 4 standard Installer s signature Date 7 Existence of other systems nearby Ye...

Page 10: ...modifications without prior notice That is why technical specifications and the products physical configuration might differ from the illustrations Make sure you get the latest news and updates about...

Reviews: