Microflow-S pocket Doppler
User’s Documentation
20 174 C
Page 4/12
2
Models & accessories
08 540 Module MICROFLOW S
Detachable component and applied part:
13 496 CWD 4MHz non preamplied probe L1.0m
Medisnap6
Detachable component and applied part:
13 503 CWD 8MHz non preamplied probe L1.0m
Medisnap6
3
Application specification
3.1 Intended medical indication
The MICROFLOW S is intended for detection of blood flow in veins and arteries and as an aid for the
diagnosis of peripheral arterial disease and venous insufficiency
.
3.2 Intended patient population
Adult only.
3.3 Intended part of the body or type of tissue applied to or interacted with.
The detachable parts of the MICROFLOW S are applied on the patient’s skin mainly of the limbs, fingers, toes
and neck.
3.4 Intended user profile
The device must be used by or on the order of physician
3.5 Intended conditions of use
The device shall be used in the specified environmental operating conditions
3.6 Expected service life
½ hour per day
200 days/year during 5 years
½ year for probes
3.7 Contact duration on applied parts and accessible parts
Probe’s active surface.
Contact time on the applied part on the same body place: less than 1mn.
3.8 Contraindication
Never use the probe on skin surfaces with recent wounds/operative cuts.
Never allow the transducer to come in contact with body fluid
It is not intended for foetal use
It not intended for use around eyes