AT7016
User Manual
7
3.3 OPERATING INSTRUCTIONS
3.3.1 VOLTAGE CHECKING
AT7016 oscilloscope is set for 110V
±
10% , 220V
±
10% voltage. Before connecting the unit to the
mains supply, make sure that the correct voltage is being used. Incorrect mains voltage may cause
damage to the instrument.
3.3.2 BASIC OPERATION
(1) POSITION FOR CONTROL SWITCHES
CONTROL SWITCHES
POSITION
INTENSITY [3]
Center
FOCUS [4]
Center
POSITION [7] [16]
Center
VOLTS/DIV [15]
0.1V
VAR [8] [17]
Calibration position
AUTO/NORM [11]
Auto
TIME/DIV [6]
0.5mS
SLOPE +/- [10]
+
INT/EXT/LINE [12]
Int
AC
⊥
DC [18]
DC
(2) OPERATION
a. Power on [1],
b. Power lights at [2],
c. Allow a warm-up period of 5 minutes, then adjust the intensity[3],
d. Adjust focus [4] for clearest line. If unstable, adjust level [9] .
(3) Horizontal Level Adjustment :
a. Slightly turn screws [21][23] counter clockwise, but do not remove;
b. Insert a straight end screw driver to [22], and whilst observing the waveform, turn the
screw driver in order to set the waveform parallel with the horizontal line ;
c. Tighten screws [21] [23].
(4) Connect the calibration signal [5] to Y input terminal[19] with 1:1 probe
(5) Adjust positions [7] [16] to obtain waveform like Fig 3-3.