
CS-3000 RF Generator Unit Operator Manual
AtriCure, Inc.
IFU-0022.A
2021/04
Page 8 / 12
CSK Type Devices, CSK-2000 Cable and RF Generator Setup
Refer to the drawing below for the appropriate setup of the CSK devices with the
CSK-2000 cable and the CS-3000 generator.
Figure 5. CSK Type Devices, CSK-2000 Cable and RF Generator Setup
CDK Type Devices, CSK-2030, CSK-2000 Cable and RF Generator Setup
Refer to the drawing below for the appropriate setup of the CDK devices with the
CSK-2000 cable, CSK-2030 cable and the CS-3000 generator.
Figure 6A. CDK Type Devices, CSK-2000, CSK-2030 with over molded
yoke and RF Generator Setup
Figure 6B. CDK Type Devices, CSK-2000, CSK-2030 with enclosure
box and RF Generator Setup
D1 = Distal Electrode 1 = Shrouded Pin #1; P1 = Proximal Electrode 1 = Shrouded Pin #3;
D2 =
Distal Electrode 2 = Shrouded Pin #2; P2 =
Proximal Electrode 2 = Shrouded Pin #4
CAUTION
: Ensure that the EP Sensing (EKG) equipment comply with IEC
60601-2-25 for protection from high frequency surgical interference.
Back Panel Connectors
Figure 7. Generator Back Panel – Key Features
Power Switch (13)
Switch that powers the generator on and off.
AC Power Connector (14)
Connector for the AC line power cable.
Grounding Stud (15)
Used as a ground equalization for safety and testing.
Data Connector (16)
USB or Serial communication connector to a host computer for data display and
archival purposes.
Alarm Volume Control (17)
Knob for modifying the volume of the generator alarm. Rotate the knob clockwise to
increase the volume.
Chapter 2 Setup and Operation
Generator Setup and Operation
Preparing the Patient – Attaching the Indifferent, Dispersive Electrode
Prepare the patient for electrosurgery following standard protocol. Ensure patient’s
entire body, including extremities, is insulated against contact with grounded metal
parts. Closely follow instructions for the coagulation device and manufacturer direc-
tions for the indifferent, dispersive electrode.
CAUTION
: Failure to achieve good skin contact by the entire adhesive surface
of the indifferent, dispersive electrode could result in a patient burn or poor electrical
performance from the coagulation device. Note: Patient Return Electrode (e.g. Indif-
ferent, Dispersive Electrode), surface area of 21 square inches (136cm
2
), equivalent
to Covidien Valleylab
Polyhesive Reference Number E7506 is recommended.
Setting up the CS-3000 RF Generator
1. Connect the supplied power cord into back mains receptacle of the CS-3000
RF Generator.
2. Plug the CS-3000 RF Generator power cord into an outlet.
3. On the back of the CS-3000, activate the mains switch and turn the RF
Generator on.
4. Turning on the generator (switch on rear of generator) causes the system
to enter Standby mode where no measurements or settings are possible.
The nContact logo and the software version number are displayed in the
message window; the Standby LED is illuminated.
5. Press the Mode button to perform a self-test and check system functionality
before transitioning to Power Control mode. Subsequent depressing of the
Mode button will toggle between Power Control and Diagnostic Evaluation
modes.
NOTE
: If an Error is detected, the red Error LED will illuminate
,
and a Message
will display in the Graphic Display. Cycle the CS-3000 main power off and on so
the RF Generator passes through self-test. (see Chapter 6: Troubleshooting).
CS-3000 Operation in Power Control Mode
When the RF Generator enters
Power Control
mode, the initial Power is set to 4 W
and Time is set to 0 until a Coagulation Device is connected via the RF coagulation
cable to the receptacle in the generator.
Connect a coagulation device to the appropriate receptacle (blue) of the cable then
connect the cable (black) to the receptacle on the RF generator so the pre-set Pow-
er and Time values are displayed. Refer to the instruction for use of the coagulation
devices for the appropriate pre-set power and time settings.