AtriCure nContact CS-3000 Operator'S Manual Download Page 5

CS-3000 RF Generator Unit Operator Manual

AtriCure, Inc.

IFU-0022.A

2021/04

Page 5 / 12

Portable and mobile RF communications equipment should be used no closer to any part of the CS-3000 RF Generator

 

including cables, than the recommended separation 

distance calculated from the equation:

 d=6/E×√P

Where: 

• 

d is the separation in meters

• 

P is the maximum output power rating of the transmitter 

in watts (W) according to the Service

• 

E is the Compliance Test Level indicated above. 

Interference may occur in the vicinity of equipment marked with the following symbol: 

a) 

 Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio 

broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an elec

-

tromagnetic site survey should be considered. If the measured field strength in the location in which the ASU System or any of its components are used exceeds 

the applicable RF compliance level above, the ASU System should be observed to verify normal operation. If abnormal performance is observed, additional 

measures may be necessary, such as reorienting or relocating components or the entire ASU System. 

b) 

 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Recommended Separation Distances between Portable and Mobile RF Communications Equipment and the CS-3000 RF Generator

The CS-3000 RF Generator is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the 

CS-3000 RF Generator can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment 

(transmitters) and the CS-3000 RF Generator as recommended below, according to the maximum output power of the communications equipment.

 

Rated Maximum

Output Power of 

Transmitter (W)

Separation Distance According to Frequency of Transmitter (m)

150 kHz to 80 MHz

d = 1.2√P

80 MHz to 800 MHz

d = 1.2√P

800 MHz to 2.5 GHz

d = 2.3√P

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance 

‘d’

 in meters (m) can be estimated using the equation applica-

ble to the frequency of the transmitter, where 

‘P’

 is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1  At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies

.

NOTE 2  These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,

 

objects and people. 

Glossary of Terms

Electrocoagulation

Surgical procedures in which high-frequency electric current is used to coagulate tissues. 

Coagulation Electrode

The metal conductor in the coagulation device used to transmit RF energy to tissue.

Indifferent, Dispersive 

Electrode

Commonly referred to as the “return electrode” or “patient electrode” or “ground pad.”   Large surface area ground used to complete the circuit 

of the electrical current.  Usually placed on the patient’s back or thigh, the Indifferent, Dispersive Electrode is connected to the generator at 

the Indifferent Connector.

Summary of Contents for nContact CS-3000

Page 1: ...restricts this device to sale by or on the order of a physician Disclaimer AtriCure reserves the right to change its products and services at any time to incorporate the latest technological developments This Manual is subject to change without notice Although this Manual has been prepared with every precaution to ensure accuracy AtriCure assumes no liability for damages resulting from the applica...

Page 2: ... and Operation 8 Generator Setup and Operation 8 Chapter 3 Cleaning 9 Chapter 4 Technical Specifications and Safety Inspection 9 Device Specifications 9 Environmental Specifications 9 Periodic Inspections 9 Chapter 5 Product Specifications 10 Operational Conditions Front Panel Displays 10 Figure 8 RF Generator Front Panel Display at Power On 10 Figure 9 RF Generator Front Panel Display at Standby ...

Page 3: ...connect to earthed ground Follow the manufacturer s directions for the placement of the indifferent dispersive electrode and for proper insulation between the patient and any metallic surfaces o Maintain safe handling techniques during electrocoagulation due to electric fields and hot metallic surfaces o Do not touch the electrode surface of the Coagulation Device and the Indifferent Dispersive El...

Page 4: ...ality should be that of a typical commercial or hospital environment Voltage Dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 Voltage Dips 0 Vt 0 5 cycle At 0 45 90 135 180 225 270 and 315 phase angles 0 Vt 1 cycle and 70 Vt 25 30 cycles Single phase at 0 Voltage interruptions 0 Vt 250 300 cycle Voltage Dips 0 Vt 0 5 cycle At 0 45 90 135 180 225 270 and 31...

Page 5: ...radiated RF disturbances are controlled The customer or the user of the CS 3000 RF Generator can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the CS 3000 RF Generator as recommended below according to the maximum output power of the communications equipment Rated Maximum Output Power of Transmit...

Page 6: ...e generator operates in either the Power Control or Diagnostic Evaluation mode When operating in Power Control mode set the desired duration and Power level The generator will transmit Power at the set point for the Time set by the operator Power may be adjusted manually throughout the treatment to tailor the coagulation process but caution should be used when deviating from the recom mended pre s...

Page 7: ...in impedance Impedance 5 Impedance resistance between the coagulation device and the indifferent dispersive electrode is measured by the generator and displayed in ohms Ω A bar graph shows the change in impedance between the initiation of RF energy baseline and throughout tissue coagulation test As conductivity decreases impedance increases During coagulation of tissue when the temperature of the ...

Page 8: ...e coagulation device and manufacturer direc tions for the indifferent dispersive electrode CAUTION Failure to achieve good skin contact by the entire adhesive surface of the indifferent dispersive electrode could result in a patient burn or poor electrical performance from the coagulation device Note Patient Return Electrode e g Indif ferent Dispersive Electrode surface area of 21 square inches 13...

Page 9: ...G Ensure Isopropyl Alcohol IPA is completely dry before oper ating the unit CAUTION Avoid caustic or abrasive cleaners Guidelines The following guidelines are recommended for cleaning the unit It is the user s responsibility to qualify any deviations from these processing methods 1 Disconnect the unit or cart from the outlet before cleaning 2 If the unit and or accessories are contaminated with bl...

Page 10: ...nd transmit RF energy to the coagulation device The RF ON OFF LED is activated Power is preset for the coagulation devices The cycle Time e g 56 s is displayed in lower right corner and counts up to the set value Impedance is measured and the value displayed in the lower left e g 105 Ω Impedance is graphed in the right Test bar against the baseline value mea sured at initiation of RF The baseline ...

Page 11: ...rected Once the warning is corrected then the parameters reset to the preset values Final Impedance 765 Ω is displayed and graphed against the baseline 100 Ω Final Percent change in impedance is displayed 665 Final Power transmitted shown 50 W Time at User Termination of RF energy 34 s displayed When this warning occurs an audible tone will sound three times for 1 5 seconds with 450 milliseconds b...

Page 12: ...ng the use of the nContact Coagulation system please contact Customer Service at AtriCure Incorporated 7555 Innovation Way Mason Ohio 45040 USA Customer Service 1 866 349 2342 toll free 1 513 755 4100 phone EC REP AtriCure Europe B V De Entree 260 1101 EE Amsterdam The Netherlands 31 20 7005560 ear atricure com WARRANTIES Limitation on Liability This warranty and the rights and obligations hereund...

Reviews: