AtriCure nContact CS-3000 Operator'S Manual Download Page 10

CS-3000 RF Generator Unit Operator Manual

AtriCure, Inc.

IFU-0022.A

2021/04

Page 10 / 12

Chapter 5 Product Specifications

Operational Conditions & Front Panel Displays

Figure 8.  RF Generator Front Panel Display at “Power On”

Turning POWER ON to the generator transitions the generator to “Standby” mode.  

Figure 9.  RF Generator Front Panel Display at “Standby Mode” 

Entering STANDBY MODE activates the “Standby” LED (“Error” LED and “RF ON/

OFF” LEDs remain off).  The nContact logo & “software version” are displayed.  

Figure 10.  RF Generator Front Panel Display Entering “Power Control 

Mode”

From Standby, when the user presses the “MODE” button, the generator passes a 

self-test.  If the self-diagnostic tests don’t detect an Error code, the generator then 

enters POWER CONTROL MODE (in this mode, the RF ON/OFF LED is blinking).

As displayed in the drawing above, the impedance measurement is Open Circuit 

(OC) because the device and the indifferent electrode is not connected.  The 

generator Power is set to 4 W and Time is set to 0.  

Figure 11.  RF Generator Front Panel Display in “Power Control \ 

Ready State”

From POWER CONTROL MODE the generator enters a 

Ready State. 

Note: An indifferent electrode and coagulation device are connected to the 

generator and placed on the patient.

“RF ON/OFF” LED is not activated, but will be blinking. 

When the user attaches a device, the generator 

detects

 device type and 

presets

 

information. 

Refer to the Instructions for Use of the coagulation devices for the pre-set power 

and time.

Figure 12. Sample Display in “RF ON” State with a Device Attached

POWER CONTROL \ 

RF ON State

 for Devices

In this mode, the user presses “

RF ON/OFF

” to activate and transmit RF energy 

to the coagulation device.  The “RF ON/OFF” LED is activated.  

Power

 is preset 

for the coagulation devices.  The cycle 

Time 

(e.g. 56 s) is displayed in lower right 

corner and counts up to the set value.

Impedance

 is measured and the value displayed in the lower left (e.g. 105 Ω).  

Impedance is graphed (in the right “Test” bar) against the baseline value mea-

sured at initiation of RF.  The baseline value is also displayed above the graph 

and in the left “Baseline” bar.

The percent change in impedance from baseline is displayed on the right (e.g.  

↓25%) with an arrow    (↑ or↓) to indicate whether the change in value was an 

increase or a decrease.

During RF application, an audible tone will sound every second for a 200 millisec-

ond pulse.

Figure 13.  Front Panel Display in “RF ON” with Resistivity Active

POWER CONTROL \ 

RF ON

 with 

Resistivity Active, Device & Indifferent, 

Dispersive Electrode Attached

“RF ON/OFF” is transmitted to the device and the “RF ON/OFF” LED is activated.

Measured impedance

 is displayed (120) but not graphed. 

Resistivity is measured

 and graphed (“Test”) against the “Baseline” value (300 Ω/cm).

Percent change in resistivity

 is displayed on the right (e.g. ↑ 20%)

Power

 transmitted at 50 W

Time

 (33 sec) counts up to set point

 

Figure 14.  RF Generator Cycle Complete, Returns to Power Control 

“Ready” State

When the RF cycle is complete upon reaching preset time, generator returns to 

the POWER CONTROL MODE 

Ready State.

“RF ON/OFF” terminates 

when Time equals set point, and LED turns off.

“Time Limit” message is displayed in the upper right.

Parameters display for 3 seconds, then reset to preset values.

Final Impedance

 is measured (80 Ω) and displayed in lower left and is graphed 

against the “baseline” value (100 Ω).

Final Percent

 change in impedance is displayed (↓ 20%).

Final Power

 transmitted shown (50W).

Time at Termination of RF energy

 is displayed (90 s).

Summary of Contents for nContact CS-3000

Page 1: ...restricts this device to sale by or on the order of a physician Disclaimer AtriCure reserves the right to change its products and services at any time to incorporate the latest technological developments This Manual is subject to change without notice Although this Manual has been prepared with every precaution to ensure accuracy AtriCure assumes no liability for damages resulting from the applica...

Page 2: ... and Operation 8 Generator Setup and Operation 8 Chapter 3 Cleaning 9 Chapter 4 Technical Specifications and Safety Inspection 9 Device Specifications 9 Environmental Specifications 9 Periodic Inspections 9 Chapter 5 Product Specifications 10 Operational Conditions Front Panel Displays 10 Figure 8 RF Generator Front Panel Display at Power On 10 Figure 9 RF Generator Front Panel Display at Standby ...

Page 3: ...connect to earthed ground Follow the manufacturer s directions for the placement of the indifferent dispersive electrode and for proper insulation between the patient and any metallic surfaces o Maintain safe handling techniques during electrocoagulation due to electric fields and hot metallic surfaces o Do not touch the electrode surface of the Coagulation Device and the Indifferent Dispersive El...

Page 4: ...ality should be that of a typical commercial or hospital environment Voltage Dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 Voltage Dips 0 Vt 0 5 cycle At 0 45 90 135 180 225 270 and 315 phase angles 0 Vt 1 cycle and 70 Vt 25 30 cycles Single phase at 0 Voltage interruptions 0 Vt 250 300 cycle Voltage Dips 0 Vt 0 5 cycle At 0 45 90 135 180 225 270 and 31...

Page 5: ...radiated RF disturbances are controlled The customer or the user of the CS 3000 RF Generator can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the CS 3000 RF Generator as recommended below according to the maximum output power of the communications equipment Rated Maximum Output Power of Transmit...

Page 6: ...e generator operates in either the Power Control or Diagnostic Evaluation mode When operating in Power Control mode set the desired duration and Power level The generator will transmit Power at the set point for the Time set by the operator Power may be adjusted manually throughout the treatment to tailor the coagulation process but caution should be used when deviating from the recom mended pre s...

Page 7: ...in impedance Impedance 5 Impedance resistance between the coagulation device and the indifferent dispersive electrode is measured by the generator and displayed in ohms Ω A bar graph shows the change in impedance between the initiation of RF energy baseline and throughout tissue coagulation test As conductivity decreases impedance increases During coagulation of tissue when the temperature of the ...

Page 8: ...e coagulation device and manufacturer direc tions for the indifferent dispersive electrode CAUTION Failure to achieve good skin contact by the entire adhesive surface of the indifferent dispersive electrode could result in a patient burn or poor electrical performance from the coagulation device Note Patient Return Electrode e g Indif ferent Dispersive Electrode surface area of 21 square inches 13...

Page 9: ...G Ensure Isopropyl Alcohol IPA is completely dry before oper ating the unit CAUTION Avoid caustic or abrasive cleaners Guidelines The following guidelines are recommended for cleaning the unit It is the user s responsibility to qualify any deviations from these processing methods 1 Disconnect the unit or cart from the outlet before cleaning 2 If the unit and or accessories are contaminated with bl...

Page 10: ...nd transmit RF energy to the coagulation device The RF ON OFF LED is activated Power is preset for the coagulation devices The cycle Time e g 56 s is displayed in lower right corner and counts up to the set value Impedance is measured and the value displayed in the lower left e g 105 Ω Impedance is graphed in the right Test bar against the baseline value mea sured at initiation of RF The baseline ...

Page 11: ...rected Once the warning is corrected then the parameters reset to the preset values Final Impedance 765 Ω is displayed and graphed against the baseline 100 Ω Final Percent change in impedance is displayed 665 Final Power transmitted shown 50 W Time at User Termination of RF energy 34 s displayed When this warning occurs an audible tone will sound three times for 1 5 seconds with 450 milliseconds b...

Page 12: ...ng the use of the nContact Coagulation system please contact Customer Service at AtriCure Incorporated 7555 Innovation Way Mason Ohio 45040 USA Customer Service 1 866 349 2342 toll free 1 513 755 4100 phone EC REP AtriCure Europe B V De Entree 260 1101 EE Amsterdam The Netherlands 31 20 7005560 ear atricure com WARRANTIES Limitation on Liability This warranty and the rights and obligations hereund...

Reviews: