23
MAINTENANCE
Maintenance Inspection
Inspection Checklist
6-6.
If any trouble is detected in any inspection, indicate on the unit that it is out of order and seek repair
without fail.
Contact your local Atom representative for further information on repairs.
WARNING
Unit name
InfaLite
Date of inspection
Serial No.
Inspector
Item to check
Results
Inspection before use (every time)
Is the main unit free of damage?
YES / NO
Is the AC adapter free of damage or deformation? Is the cable of AC adapter free of damage or
deformation?
YES / NO
Is the cable cover free of damage?
YES / NO
Does pressing the stand-by switch and activating the non-contact sensor cause its light to turn on? YES / NO
Does pressing the stand-by switch and holding the palm of your hand within 15cm of the non-contact
sensor cause the sensor to respond and the light to turn on?
YES / NO
Does turning on the power, turning on the light with the non-contact sensor, then pressing “
−
” or
“
+
” on the light control button cause the light intensity to change?
YES / NO
Is the lens cover free of any cracks, dirt, or damage?
YES / NO
Is the flexing arm free of damage? Is the irradiation head held in place?
YES / NO
Is the linking arm free of damage? Do the movable parts rotate?
YES / NO
Is the F-Rail Attachment/Pole Mounting Attachment (sold separately) securely attached to the
incubator, etc.?
YES / NO
Is the Operation Manual kept where it is readily accessible?
YES / NO
Monthly inspection
Is the F-Rail Attachment/Pole Mounting Attachment (sold separately) securely attached to the incu-
bator, etc.? Can the F-Rail Attachment/Pole Mounting Attachment be detached from the incubator,
etc. by loosening the fixing knob?
YES / NO
Are the parts of the unit deteriorated or damaged (cracked, bent, damaged, etc.)?
YES / NO
Does the white LED emit light?
YES / NO
Does the flexing arm hold the irradiation head in place?
YES / NO
Does each movable part of the linking arm rotate without becoming loose?
YES / NO
Does the stand-by indicator on the irradiation head turn “green” when the AC adapter plug is in-
serted into the power outlet?
YES / NO
Semiannually inspection
Does the light intensity of the white LED reach the specified value (see “6-1. Measuring the Light
Intensity”)?
YES / NO