2. If LOAD1 is bigger than 80% capacity, such as full capacity, after first load calibration, it will

finish automatically, you only need to press

TARE

to move decimal position. Generally, we

strong recommend you to use this one point calibration program if you have 100% capacity

standard weight

3. If you only have 20% standard weight, after first load calibration, when it display

LoAd2

,

press

OFF

2 times, then press

TARE

to move decimal position. Calibration finish.

CHAPTER 5 DISPLAY ILLUSTRATION

DISPLAY

ILLUSTRATION

REMARK

CALSP

Enter zero point calibration

SCALE

Enter calibration

SEtUP

Enter parameter setting

UAdJ

Enter voltage calibration

LoAd1

First calibration point

LoAd1<LoAd2<LoAd3

LoAd2

Second calibration point

LoAd3

Third calibration point

¯¯¯¯¯

Exceed high limit

Tare weight can’t exceed full capacity

_____

Exceed low limit

Tare weight can’t be negative

-----

Waiting

Err10

Weight less than Min. Capacity

Can’t accumulate the value

Err11

Accumulated times overflow

Can’t accumulate after 30times

Err12

Accumulated weight overflow

Can’t accumulate after 99999

Err13

Error in repeat accumulation

Can’t accumulate one weight repeatedly

no***

Current accumulation times

H****

Front four digit of accumulated
weight

Total weight=front four digit + rare four digit

L****

Rare four digit of accumulated
weight

Total weight=front four digit + rare four digit

CLr

Ask if you really want to delete
accumulated weight

In case error deletion

noCLr

Give up deletion

88888

Confirm deletion

SHIFt

Switch

¯¯¯

Input value is too large

When you input tare or weight value

___

Input value is too small

When you input tare or weight value

noACC

No any accumulated content

when you check accumulation

-oL-

Overload warning

Tare + Net weight exceed full ca 9e

-Lb-

Low battery warning

Turn off automatically one minute later

Digital Crane Scale

1

2

Automation       www.ato.com      [email protected]     Global Shipping      +1 800-585-1519 (Toll-free)

Summary of Contents for GSP911-30

Page 1: ...DIGITAL CRANE SCALE USER MANUAL SKU ATO CRS 50T Digital Crane Scale Automation www ato com sales ato com Global Shipping 1 800 585 1519 Toll free...

Page 2: ...ontrol 5 CHAPTER 3 OPERATION GUIDE 6 1 Turn on off 6 2 Zero 7 3 Tare 7 4 Hold 7 5 Accumulation 7 6 Accumulation Search 7 7 Accumulation Clear 8 8 Battery Voltage 8 9 Brightness 8 10 Zero Point 8 11 Un...

Page 3: ...ery every three months to preserve the life span of the battery and recharge it before operation 5 The battery charger is included in the scale package Please use this charger It s normal that the cha...

Page 4: ...Battery Operated with low battery indication Automatic turn off power saving function 2 Main technical data Standard According to OIML class III Display 45mm 1 8 5digits LED Zero Range 4 F S Tare Ran...

Page 5: ...it can be used for about 70hours Low battery warning When the digits flash it means the battery needs to be recharged The scale will power off automatically after 1 to 2 hours operation if the batter...

Page 6: ...2 Put the goods into container the scale will display goods net weight 3 If move goods and container the scale will display minus value of tare weight 4 REMOTE CONTROLLER Press any key on the remote...

Page 7: ...ightness of LED 7 F2 Function Key used with F1 together 8 HOLD It used for hold function details see Chapter3 4 9 OFF Turn off the crane scale CHAPTER 3 OPERATION GUIDE express key on remote controlle...

Page 8: ...n goods it will display net weight of goods 4 HOLD OPERATION ILLUSTRATION Press HOLD Hold the weight reading no matter load change Press HOLD again Return to normal weighing status 5 ACCUMULATION OPER...

Page 9: ...should be between 5 8 and 6 9 if below than 5 8 the digits flash you need to recharge immediately 9 BRIGHTNESS OPERATION ILLUSTRATION F1 1 5 total 5 options from darkness to brightness Default is 3 1...

Page 10: ...take cap 30000kg as example choose FS 30 Press TARE confirm Id 05 New scale the screen display Id 05 if scale calibrated the screen display calibrated division Press ZERO or CLEAR to choose Id 10 Choo...

Page 11: ...if scale calibrated the screen display calibrated value Press ZERO or CLEAR to choose OFF 2 Choose turn off method 0 2 total 3 options 0 turn off by hand 1 if no operation the screen display after 15...

Page 12: ...g 20000kg Second load must exceed Load 1 20 capacity Waiting until STB light on Press SHIFT 16000 Mention you Load2 must exceed 16000kg Press TARE ZERO CLEAR 20000 Input actual weight 20000kg waiting...

Page 13: ...t be negative Waiting Err10 Weight less than Min Capacity Can t accumulate the value Err11 Accumulated times overflow Can t accumulate after 30times Err12 Accumulated weight overflow Can t accumulate...

Page 14: ...or PCB Requires authorized service Load cell cable damaged Check and replace load cell Out of calibration Check calibration Display experiences excessive Zero drift between weighment Scale electronic...

Reviews: