2. calibration

OPERATION

DISPLAY

ILLUSTRATION

0

Take Cap:30000kg scale as example

Pay attention that

the screen display 0 after hanging tare weight, you can
hang tare weight at first, then turn on.

Press

F1

and

F2

at

same

time

3seconds

P0000

Second digit flash

use function key to move digit, set

value.

Press

TARE

ZERO

】【

CLEAR

P1358

Input 1358

1358 is calibration password

Press

SHIFT

SCALE

Confirm

into calibration program

Press

SHIFT

CALSP

Zero point calibration, wait until STB light on

Press

SHIFT

LoAd1

First load calibration

Hanging on first load
e.g.10000kg

First load must exceed 20% capacity, Waiting until STB
light on.

Press

SHIFT

50000

Display capacity

Press

TARE

ZERO

】【

CLEAR

10000

Input 10000kg, wait until STB light on.

Press

SHIFT

LoAd2

Second load calibration

Hanging on second
load e.g.20000kg

Second load must exceed Load 1+20% capacity,
Waiting until STB light on.

Press

SHIFT

16000

Mention you Load2 must exceed 16000kg

Press

TARE

ZERO

】【

CLEAR

20000

Input actual weight 20000kg, waiting stable

Press

SHIFT

LoAd3

Third load calibration

Hanging on third load
e.g.30000kg

Third load must exceed Load 2+20% capacity, Waiting
until STB light on.

Press

SHIFT

26000

Mention you Load3 must exceed 26000kg .

Press

TARE

ZERO

】【

CLEAR

30000

Input actual weight 30000kg. wait until STB light on.

Press

SHIFT

------ -

Waiting stable

-oL-

End

Calibration end

30000

Display calibrated capacity , sometimes it display
0000.5, 9999.5

Press

TARE

30000

Move decimal position

Put down weight

0

Check zero point, it display 0, and STB, ZERO light on.

1. First load: LOAD1 must larger than 20% capacity

and LOAD1<LOAD2<LOAD3, at the same time,

between LOAD1 and LOAD2

LOAD2 and LOAD3, the interval must bigger than 20% capacity

Digital Crane Scale

1

1

Automation       www.ato.com      [email protected]     Global Shipping      +1 800-585-1519 (Toll-free)

Summary of Contents for GSP911-30

Page 1: ...DIGITAL CRANE SCALE USER MANUAL SKU ATO CRS 50T Digital Crane Scale Automation www ato com sales ato com Global Shipping 1 800 585 1519 Toll free...

Page 2: ...ontrol 5 CHAPTER 3 OPERATION GUIDE 6 1 Turn on off 6 2 Zero 7 3 Tare 7 4 Hold 7 5 Accumulation 7 6 Accumulation Search 7 7 Accumulation Clear 8 8 Battery Voltage 8 9 Brightness 8 10 Zero Point 8 11 Un...

Page 3: ...ery every three months to preserve the life span of the battery and recharge it before operation 5 The battery charger is included in the scale package Please use this charger It s normal that the cha...

Page 4: ...Battery Operated with low battery indication Automatic turn off power saving function 2 Main technical data Standard According to OIML class III Display 45mm 1 8 5digits LED Zero Range 4 F S Tare Ran...

Page 5: ...it can be used for about 70hours Low battery warning When the digits flash it means the battery needs to be recharged The scale will power off automatically after 1 to 2 hours operation if the batter...

Page 6: ...2 Put the goods into container the scale will display goods net weight 3 If move goods and container the scale will display minus value of tare weight 4 REMOTE CONTROLLER Press any key on the remote...

Page 7: ...ightness of LED 7 F2 Function Key used with F1 together 8 HOLD It used for hold function details see Chapter3 4 9 OFF Turn off the crane scale CHAPTER 3 OPERATION GUIDE express key on remote controlle...

Page 8: ...n goods it will display net weight of goods 4 HOLD OPERATION ILLUSTRATION Press HOLD Hold the weight reading no matter load change Press HOLD again Return to normal weighing status 5 ACCUMULATION OPER...

Page 9: ...should be between 5 8 and 6 9 if below than 5 8 the digits flash you need to recharge immediately 9 BRIGHTNESS OPERATION ILLUSTRATION F1 1 5 total 5 options from darkness to brightness Default is 3 1...

Page 10: ...take cap 30000kg as example choose FS 30 Press TARE confirm Id 05 New scale the screen display Id 05 if scale calibrated the screen display calibrated division Press ZERO or CLEAR to choose Id 10 Choo...

Page 11: ...if scale calibrated the screen display calibrated value Press ZERO or CLEAR to choose OFF 2 Choose turn off method 0 2 total 3 options 0 turn off by hand 1 if no operation the screen display after 15...

Page 12: ...g 20000kg Second load must exceed Load 1 20 capacity Waiting until STB light on Press SHIFT 16000 Mention you Load2 must exceed 16000kg Press TARE ZERO CLEAR 20000 Input actual weight 20000kg waiting...

Page 13: ...t be negative Waiting Err10 Weight less than Min Capacity Can t accumulate the value Err11 Accumulated times overflow Can t accumulate after 30times Err12 Accumulated weight overflow Can t accumulate...

Page 14: ...or PCB Requires authorized service Load cell cable damaged Check and replace load cell Out of calibration Check calibration Display experiences excessive Zero drift between weighment Scale electronic...

Reviews: