2. calibration
OPERATION
DISPLAY
ILLUSTRATION
〖
0
〗
Take Cap:30000kg scale as example
,
Pay attention that
the screen display 0 after hanging tare weight, you can
hang tare weight at first, then turn on.
Press
【
F1
】
and
【
F2
】
at
same
time
3seconds
〖
P0000
〗
Second digit flash
,
use function key to move digit, set
value.
Press
【
TARE
】
【
ZERO
】【
CLEAR
】
〖
P1358
〗
Input 1358
,
1358 is calibration password
Press
【
SHIFT
】
〖
SCALE
〗
Confirm
,
into calibration program
Press
【
SHIFT
】
〖
CALSP
〗
Zero point calibration, wait until STB light on
Press
【
SHIFT
】
〖
LoAd1
〗
First load calibration
Hanging on first load
e.g.10000kg
First load must exceed 20% capacity, Waiting until STB
light on.
Press
【
SHIFT
】
〖
50000
〗
Display capacity
Press
【
TARE
】
【
ZERO
】【
CLEAR
】
〖
10000
〗
Input 10000kg, wait until STB light on.
Press
【
SHIFT
】
〖
LoAd2
〗
Second load calibration
Hanging on second
load e.g.20000kg
Second load must exceed Load 1+20% capacity,
Waiting until STB light on.
Press
【
SHIFT
】
〖
16000
〗
Mention you Load2 must exceed 16000kg
Press
【
TARE
】
【
ZERO
】【
CLEAR
】
〖
20000
〗
Input actual weight 20000kg, waiting stable
Press
【
SHIFT
】
〖
LoAd3
〗
Third load calibration
Hanging on third load
e.g.30000kg
Third load must exceed Load 2+20% capacity, Waiting
until STB light on.
Press
【
SHIFT
】
〖
26000
〗
Mention you Load3 must exceed 26000kg .
Press
【
TARE
】
【
ZERO
】【
CLEAR
】
〖
30000
〗
Input actual weight 30000kg. wait until STB light on.
Press
【
SHIFT
】
〖
------ -
〗
Waiting stable
〖
-oL-
〗
〖
End
〗
Calibration end
〖
30000
〗
Display calibrated capacity , sometimes it display
0000.5, 9999.5
Press
【
TARE
】
〖
30000
〗
Move decimal position
Put down weight
〖
0
〗
Check zero point, it display 0, and STB, ZERO light on.
1. First load: LOAD1 must larger than 20% capacity
,
and LOAD1<LOAD2<LOAD3, at the same time,
between LOAD1 and LOAD2
,
LOAD2 and LOAD3, the interval must bigger than 20% capacity
Digital Crane Scale
1
1