Patient leakage current test
Measurement:
See User’s Handbook of the respective safety tester (patient leakage current test).
Adapt a mains cable for the device and the safety tester and take the device into service.
The patient leakage current is measured on all patient connections/signal inputs and signal outlets that are connected together
(users parts).
Record patient leakage current before and after pole changing in the test report.
Measuring points:
Overfl ow contacts on the sides
Overfl ow contacts on the housing fl oor
Permissible measured value:
N.C. < 10
µ
A
The following errors are checked:
•
Mains interrupted: press mains SFC key, read value and record in test report.
Permissible measured value:
S.F.C. < 50
µ
A
•
Protective earth conductor: Press SFC SL key, read value and record in test report.
Permissible measured value:
S.F.C. < 50
µ
A
Insulation testing (high voltage)
The insulation test must not be carried out. It was already carried out before the device was delivered and should not be
repeated.