44
Display
Cause of error
Troubleshooting
If the device is in a tilted
position the warning „Device in
critical tilt” appears.
Place the device in an upright
position. The warning signal
is automatically deactivated.
If the therapy has not been
started after initial operation
the warning message "Inactive
therapy" appears” without
tone.
Press the Play symbol to start
the therapy.
If the warning message is
suppressed, it reappears
after one minute, but without
tone.
If the therapy has already
started and has been
interrupted by the pause
symbol, the warning message
"Inactive therapy" appears with
tone.
Press the Play symbol to start
the therapy.
If the warning message is
suppressed, it appears again
with tone after one minute.
The device must no longer be
operated.
Possible causes:
• Battery or
•
pump defective.
Contact the ATMOS service.
Provide sufficient ventilation!
Check fan!
Device temperature too high
WARNING
Device temperature too high.
• Device is in the sun or near a
heater.
•
Ventilation slots are covered.
•
Fan is defective.
•
Place the device in a cooler
location.
•
Please ensure sufficient air
ventilation.
•
Please contact the ATMOS
service.