9
Other contraindications:
•
No application with other cardiothoracic drainage
systems other than the ATMOS C 051 Thorax, ATMOS S
201 Thorax and ATMOS E 201 Thorax
•
No separate application of the secretion canister and
the hose system (i.e. without device) as gravity drainage.
•
No application under emergency conditions
•
Should not be used in the homecare sector which is not
supervised by healthcare professionals
•
No suction of flammable, corrosive or explosive fluids/
gases
Warnings:
The following complications may occur during a
cardiothoracic drainage:
•
Pain due to irritation of the intercostal nerves
•
Injuries to the lung parenchyma / air leak
• Re-expansion edema
•
Effusion retention
• Tension pneumothorax
•
Skin / soft tissue emphysema
The product is:
not active
Sterility / specific microbial
condition:
Hose system is sterile.
Single-use product /
reprocessing:
Hose system is a disposable product.
1.3 Function
The ATMOS S 201 Thorax is an exceptionally handy, portable, digital thoracic drainage suction
device. The device is operated with an electrical, maintenance-free diaphragm pump. During
operation the pump creates a vacuum within the suction hose and the secretion canister by
means of which secretion and air can be sucked off by the hose system. The pump is controlled
digitally and therefore ensures that the chosen required vacuum value is stable. The air flow,
which is measured in real time, is displayed in numbers. At one key press flow data from up to 12
days is shown in graphic.
The secretion is collected in the secretion canister. Its capacity is 2 l. With the aid of the
measuring and rinsing hose the vacuum at the end of the hose system is measured. Via the
touch screen display the target vacuum can be set manually. The suction capacity is regulated
automatically.
The hose system is rinsed with air at regular intervals to prevent the collection of debris in the
secretion hose. This measure also prevents secretion from intruding into the measuring and
rinsing hose or that a syphon effect is created.
The device is equipped with a rechargeable battery. A charging unit which is located within the
suction device guarantees for the secure charging of the battery. Therefore it is impossible to
overcharge the battery.
Bacterial and viral filters in the secretion canister and measuring hose prevent contaminated
secretions from entering the device. As accessories a universal connection and a carrying strap
can be ordered separately. These enable attachment to the patient's bed, for example, as well as
mobilization.
Essential requirements:
•
Create and maintain a regulated vacuum between -5 and -100 mbar.