Atmos LS 21 LED Operating Instructions Manual Download Page 4

4

1.1  Notes on Operating Instructions

 1.0  Introduction

These operating instructions contain important notes on how to operate the ATMOS

®

 LS 21 LED safely, 

correctly and effectively. Their reading helps to avoid risks, and also to reduce repair costs and down-

times. This increases, amongst other things, the reliability and service-life of the device.
These operating instructions serve not only for new operating personnel to be instructed in its use, but 

also for use as a reference manual. Reprints (also in extracts) only with permission in written form by 

ATMOS.

These operating instructions must always be kept available near the device.

•  The product ATMOS

®

 LS 21 LED bears CE marking CE according to the EC Directive of the council 

for medical products 93/42/EEC and meets the basic requirements of Appendix I of the directive.

•  The product ATMOS

®

 LS 21 LED complies with all applicable requirements of the Directive 2011/65/

EC restricting the use of certain hazardous substances in electrical and electronic equipment 

(“RoHS”).

•  The declaration of conformity and our general standard terms and conditions can be obtained on our 

website at www.atmosmed.com.

• 

The quality management system applied at ATMOS has been certified according to international 

standards EN ISO 13485.

•  Prior to start-up please peruse chapter 2.0 „For your safety“, in order to be prepared for any possible 

dangerous situations.

1.3 Function

The light source ATMOS LS 21 LED is used to connect it 

to optical instruments like rigid and flexible endoscopes 

respectively laryngoscopes for illumination of throat, nose 

and ear. 
The anodized aluminium lamp body of the ATMOS

®

 LS 21 

LED is slender which enables an ergonomic fixation to the 

instrument.
There is an adapter sleeve available for adaptation to 

current endoscopes which enables the connection of optics 

from Storz, Wolf and as per ACMI standard. The optical 

instruments with an ATMOS adapter sleeve snap in at the 

light source with a ball-shaped quick lock.

The power supply for the light source may either be an 

ATMOS power source (power source integrated in the 

ENT unit) for LED lamps or the rechargeable LED battery 

(please also see HL 21 LED headlight).

The light source has a fix connecting cable with a plug, 

which fits both to the rechargeable battery or to the power 

supply of the unit.

The light of the 2 Watt LED is beamed into the optical 

instrument with a launching lens.

The electrical components (connecting lead and LED) are 

installed in the housing and are isolated (4kV). Therefore, 

the light source fulfils the requirements for endoscopes 

respectively its light sources in accordance with DIN EN 

60601-2-18 (labelling as an application part type BF). 

1.2  Intended use

Name: 

ATMOS

®

 LS 21 LED

Main function:

LED light source for optical instruments to illuminate body 

openings (ear, nose, throat).

Medical indications / application:

Illumination of body orifices (ear, nose, throat)

Specification of the main function:

Type of light source: LED lamp (high intensity)
Energy source: Rechargeable battery or current source

Application organ:

ENT examination: Ear, nose, throat
OT: For use in the OT by ENT doctors, gynaecologists, 

cardiologists, orthopaedics, cosmetic/plastic surgeons, 

proctologists

Application time:

ENT: temporary (up to 60 mins)
OT: for a maximum of 4 hours

Application site:

ENT examination: in clinic and practice of ENT doctors
OT: Operating theatre
The light source may only be used by medical trained staff.

Contraindications:

Do not use in explosive areas.

The product is:

 Active

Sterility:

Not necessary

Single-use product / reprocessing:

No single use product

Summary of Contents for LS 21 LED

Page 1: ...ATMOS LS 21 LED Light source GA1GB 150302 0 2018 10 Index 14 Operating Instructions English ...

Page 2: ...ATMOS MedizinTechnik GmbH Co KG Ludwig Kegel Straße 16 79853 Lenzkirch Germany Phone 49 0 76 53 689 0 Fax 49 7653 689 190 49 7653 689 393 Service Center atmos atmosmed de www atmosmed de ...

Page 3: ...he application part 7 4 2 Connection possibilities of the light source 7 4 3 Operation of the optional LED battery 8 4 3 1 Indication of capacity 8 4 3 2 Reloading the LED battery 8 4 3 3 Maintenance of battery 8 5 0 Cleaning 9 6 0 Repairs and servicing 10 6 1 Sending in the device 10 7 0 Troubleshooting 10 8 0 Accessories and spare parts 11 8 1 Accessories 11 8 2 Spare parts 11 9 0 Technical data...

Page 4: ...tion to the instrument There is an adapter sleeve available for adaptation to current endoscopes which enables the connection of optics from Storz Wolf and as per ACMI standard The optical instruments with an ATMOS adapter sleeve snap in at the light source with a ball shaped quick lock The power supply for the light source may either be an ATMOS power source power source integrated in the ENT uni...

Page 5: ... correct fit of the light source in the holders Only use power sources manufactured by ATMOS Prior to starting up the light source ATMOS LS 21 LED read these operating instructions carefully Attention The light handle gets warmer during application ATMOS is not liable for personal injury and damage to property or malfunctions if no original ATMOS parts are being used the advice for use in these op...

Page 6: ...ndoscope with integrated light conducting cable Light conducting adaptor Storz REF 507 4611 0 with detent REF 507 4612 0 for screwing Endoscope adaptor Wolf Light conducting adaptor Storz Danger to the device CAUTION If the light source is used with unsuitable adaptors the lens could be badly damaged Use optics only with the supplied and or optional adaptors Make sure the endoscopes with adaptor e...

Page 7: ...plication part 4 0 Operation click Endoscopes ATMOS Cam 21 31 ATMOS LS 21 LED light source Endoscopes with light conducting cable LED battery 4 2 Connection possibilities of the light source OR Adaptor Storz Adaptor Olympus ...

Page 8: ...d when all 5 LEDs expire again 1 2 4 3 3 Maintenance of battery During storage During usage Do not Suck it out When the capacity indication shows empty all LEDs are out do not use the battery any more Recharge immediately For flawless operation the battery needs a residue capacity Expose the battery to solar radiation or put it close to heaters This will reduce the lifespan of the battery After ap...

Page 9: ...s or anionic tensides as these may cause stress cracks in the material used for the housing of the unit Do not moisten the lens in the light object with liquids The surfaces can be cleaned with a moist cloth The device has to be cleaned and disinfected regularly if it is applied by several users Surface disinfectants Disinfectant Ingredients in 100 g Manufacturer Green Clean SK Di alkyl dimethyl a...

Page 10: ...claims whatsoever on defects or malfunctions which arise from the use of third party accessories or consumables 6 1 Sending in the device Remove and properly dispose of consumables Clean and disinfect the product and accessories according to the operating instructions Place used accessories with the product Fill in the form QD 434 Delivery complaint return shipment and the respective decontaminati...

Page 11: ...rsed without light guide 950 0212 0 Ear endoscopes Wide angle optic 0 Working length 50 mm Ø 4 mm autoclavable 950 0213 0 Wide angle optic 30 Working length 50 mm Ø 4 mm autoclavable 950 0214 0 Wide angle optic 0 Working length 34 mm Ø 2 7 mm can be immersed Incl adapter for ear speculum 950 0215 0 Nose Pharynx endoscopes Ø 4 mm Wide angle optic 0 Working length 180 mm Ø 4 mm autoclavable 950 0216...

Page 12: ...K Light emitter warm white high performance LED Electricity 700 mA Performance 2 W Light intensity Min 150 klx Light flux Min 120 lm Colour temperature Type 3900 K Ambient conditions transport storage Temperature 20 50 C Humidity without condensation 5 90 Air pressure 700 1060 hPa Ambient conditions for operation Temperature 10 35 C Humidity without condensation 20 80 Air pressure 700 1060 hPa Per...

Page 13: ...hould ensure that it is used in such an environment Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment Guidance Electrostatic discharge ESD according to IEC 61000 4 2 6 kV Contact 8 kV Air 6 kV Contact 8 kV Air Floors should be made of wood or concrete or tiled with ceramic tiles If floors are synthetic the relative humidity should be at least 30 EFT IEC 61000 4 4 2 kV...

Page 14: ...vels 11 3 Guidelines and Manufacturer s Declaration Immunity The ATMOS LS 21 LED is intended for use in the electromagnetic environment specified below The customer or user of the ATMOS LS 21 LED should ensure that it is used in such an environment Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment Guidance Conducted RF IEC 61000 4 6 3 Veff 150 kHz to 80 MHz 10V Portab...

Page 15: ...80 MHz the field strength should be below 3 V m The ATMOS LS 21 LED is intended for use in electromagnetic environment in which radiated disturbances are controlled The customer or user of the ATMOS LS 21 LED can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF Communications equipment and the ATMOS LS 21 LED as recommended below according ...

Page 16: ...ATMOS MedizinTechnik GmbH Co KG Ludwig Kegel Straße 16 79853 Lenzkirch Germany Phone 49 7653 689 0 atmos atmosmed de www atmosmed com ...