11.1 Guidelines and Manufacturer's Declaration - Emissions
The ATMOS
®
LS 21 LED light source is intended for use in the electromagnetic environment specified below. The customer or
user of the ATMOS
®
LS 21 LED should ensure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF Emissions acc.to CISPR 11
Group 1
The ATMOS
®
LS 21 LED uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
RF Emissions acc. to CISPR 11
Class B
The ATMOS
®
LS 21 LED is suitable for use in all
establishments, including domestic and those directly
connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions according to
IEC 61000-3-2
Inapplicable
Voltage fluctuations/flicker
acc. to IEC 61000-3-3
Inapplicable
11.2 Guidelines and Manufacturer's Declaration - Immunity
The ATMOS
®
LS 21 LED is intended for use in the electromagnetic environment specified below. The customer or user of the
ATMOS
®
LS 21 LED should ensure that it is used in such an environment.
Immunity Test
IEC 60601- Test Level
Compliance Level
Electromagnetic Environment - Guidance
Electrostatic discharge
(ESD) according to IEC
61000-4-2
± 6 kV Contact
± 8 kV Air
± 6 kV Contact
± 8 kV Air
Floors should be made of wood or concrete
or tiled with ceramic tiles. If floors are
synthetic, the relative humidity should be at
least 30 %.
EFT IEC 61000-4-4
± 2 kV Mains
± 1 kV I/Os
± 2 kV Mains
Inapplicable
The quality of the supply voltage should
correspond to a typical commercial or
hospital environment.
Surges IEC 61000-4-5
1 kV
Common
2 kV
Differential
1 kV
symmetric
Inapplicable
The quality of the supply voltage should
correspond to a typical commercial or
hospital environment.
Magnetic field at power
frequency 50/60 Hz acc.
to IEC 61000-4-8
3 A/m
Inapplicable
Power frequency magnetic fields should
be that of a typical commercial or hospital
environment.
• Medical electrical equipment is subject to special precautions with regard to EMC and must be installed acc. to following
EMC notes.
•
Portable and mobile HF communication facilities can influence medical electrical equipment.
• The use of other accessories, other transducers and cables than stated may lead to an increased emission or a reduced
interference immunity of the equipment or system.
11.0 Notes on EMC