3.0 Setting up and starting
3.1
Installation in the auxiliary
housing
The device is mounted in a shapely housing next to the
treatment unit on a level and suffi ciently sustainable
surface.
Main switch:
It must be ensured that the power supply is
connected downstream of the main instrument /
practice switch.
Connect the device to a properly installed earthed contact
socket. Adhere to the connected load stated on the type
plate.
In case of unfi rm assembly, there is a danger of liquid
from the mixing container’s overfl ow spilling out and
activating the overfl ow sensor (see chapter 4.2).
Starting Sequence:
1. Open main water tap
2. Switch on ATMOS S 61 Servant
3. Switch on ATMOS Aqua clean
If error “F1” appears, switch the ATMOS S 61 Servant
off and on, until the error disappears.
3.2
Hose connections
Chemical bottle connection
PVC tubing
Ø internal = 2 mm
Ø outer = 4 mm
Connection to the water outlet
(supplying the instruments)
PVC tubing
Ø internal = 4 mm
Ø outer = 6 mm
Water connection (fresh water in drinking water
quality)
Permitted pressure at entry: 2-6 bar
PVC tubing (with KTW authorisation)
Ø internal = 6 mm
Ø outer = 8 mm
3.2.1 Initial connection of the
chemical bottle
Open the housing of the ATMOS Aqua clean and feed the
hose (Fig. 4) through the housing to the outside. Plug
the provided seal onto the bottle and insert the hose. Fix
the bottle in the provided support.
Fig. 3
1