6.1 Preparatory measures
The following steps are necessary to start up the equip-
ment combination of the SunCool instrument with the
SUNTEST instrument:
1. Remove the filter mat from the air inlet aperture on the
SUNTEST instrument.
2. Insulate the housing surface of the SUNTEST CPS+
instrument (optional).
3. Connect the drain hose to the evaporator unit.
Connect the micro-pump for the waste water to the
connection nozzle of the evaporator unit optionally.
4. Install the liquefier unit. Install the liquefier unit optionally
outside and position the electric cable for the power
supply as well as the connecting cables between the
liquefier unit and the evaporator unit.
5. Install the SUNTEST CPS+ instrument and place the
evaporator unit on the SUNTEST instrument.
6. Connect the liquefier unit and the SUNTEST instrument
using the interface cable.
7. Connect the drain hose of the evaporator unit to the
waste water tank or sewer drain.
8. Connect the drain hose of the micro-pump to the waste
water tank or sewer drain.
9. Connect the liquefier unit to the power supply.
10. Connect the SUNTEST instrument to the power supply
and start.
Insulating the surfaces of the SUNTEST instrument
(Fig. 5)
The three insulating pads prevent condensation from
forming on the housing surfaces in the areas:
• air inlet aperture
2
of the SUNTEST CPS+
1
NOTE – adhesion:
The insulating pads are strongly adhesive and, once
applied, are very difficult to remove.
Therefore, mark the position of the insulating pads before
applying them to the housing surfaces.
Attaching insulating pads:
1. Clean the housing surface with a commercially available
cleaning agent. The adhesion areas must be dry and
free from grease and dust.
2. Mark the position on the right and rear sides of the
instrument and on the inside top of the housing, pull
off the backing foil, position the insulating pads and
press firmly.
Fig. 5
6
SunCool start up
1
2