Operating Manual SunCool for SUNTEST
®
CPS+
3
Requirements for the installation site
3.1 Room climate
Climatic requirements for the operating room:
In continuous operation together with the SUNTEST CPS+
there is a constant change in the room climate due to the
hot air emitted by the radiator cooling and the liquefier
unit as well as the cold air emitted by the evaporator unit.
Therefore, the device may only be installed in adequately
aired, dust-free rooms or must be connected to a suitable
technical ventilation unit.
Ambient conditions:
• Room temperatures of 18 °C to 25 °C
• Relative humidity of 50% max.
3.2 Room ventilation
The coolant is processed in the cooling system in different
aggregate states. The coolant is heavier than air in vapor
form. The room must therefore have sufficient ventilation
facilities near the floor. The doors of the room must have
ventilation slits near to the floor.
3.3 Installation
• Evaporator dimensions (W x D x H):
35 cm x 32 cm x 33 cm
• Evaporator weight: approx. 7 kg
• Liquefier unit dimensions (W x D x H):
70 cm x 60 cm x 35 cm
• Liquefier unit weight: 30 kg
• Place the evaporator unit on a stable laboratory bench
with a non-flammable surface and align the lab bench
horizontally
NOTE –
minimum load capacity of the lab bench
The total dead weight of the SUNTEST
®
CPS+ and a
SunCool is approx. 40 kg. Please note the minimum load
capacity of your lab bench.
CAUTION – overheating of the device!
If the air feed and discharge (of the SUNTEST CPS+) is
disturbed, the device overheats and can be damaged!
Keep the following minimum distances from adjacent
walls or objects when installing:
Front: 70 cm
Rear: 50 cm
Top:
70 cm
Sides: 50 cm