![Atlantic Ultraviolet Nutripure 2B-SC Owner'S Manual Download Page 8](http://html.mh-extra.com/html/atlantic-ultraviolet/nutripure-2b-sc/nutripure-2b-sc_owners-manual_3000441008.webp)
8
t
eChniCal
a
ssistanCe
Atlantic Ultraviolet Corporation
®
maintains a qualified team of application specialists that can provide technical assistance, which
may be required during the use of this equipment. For any questions, concerns, or technical assistance, regarding this model or
any of our other products, contact our application specialists. Contact options can be found in the “
User Assistance
” section on
the back page of this manual.
MAINTENANCE
Always disconnect power to the
Nutripure
®
before performing any service or maintenance.
•
The
Nutripure
®
is designed to operate with a minimal amount of maintenance.
•
A regular cleaning cycle of the
STER-L-RAY
®
Germicidal Ultraviolet Lamp(s), interior,
Promate
TM
sight glass and intake
filter(s)
MUST
be established, and carefully maintained, based on inspection and experience. The frequency of cleaning will
vary with the conditions surrounding each installation. It is recommended that the ultraviolet lamp(s) interior, sight glass, and
filters be cleaned at least once every three (3) months.
•
The
STER-L-RAY
®
Germicidal Ultraviolet lamp(s) used in the
Nutripure
®
have a manufacturer’s rated average effective
life of 10,000 hours.
STER-L-RAY
®
Lamp(s) may operate longer than their rated effective life, but the reduction
in ultraviolet output will make it impractical to use past the manufacturer’s rated life. For maximum efficiency lamp
replacement is recommended every 10,000 hours of operation or about one (1) year of continuous use.
•
Unless lamp replacement is due to failure or breakage, it is recommended that all lamps be replaced at the same time.
l
amP
i
nstallation
, r
ePlaCement
,
or
C
leaning
IN ORDER TO PERFORM THIS TASK, BE SURE TO WEAR THE FOLLOWING SAFETY
EQUIPMENT: SAFETY GLASSES OR A FACE SHIELD, AS WELL AS GLOVES.
Always disconnect power to the fixture before performing any service or maintenance.
Ultraviolet lamps are easily damaged. Exercise care when handling lamps.
1. Disconnect power to the sanitary conditioner and unplug from the electrical outlet.
2. Remove access cover. (Removal of the
Nutripure
®
access cover will cause the interlock switch to interrupt power to the
lamps and blower.
DO NOT
attempt to bypass or circumvent the interlock switch or its safety feature.)
3. Remove
STER-L-RAY
®
lamp(s), carefully press lamp against the spring-loaded lampholder, when
STER-L-RAY
®
lamp
pin clears the stationary lampholder, pull
STER-L-RAY
®
lamp out, and away from
Nutripure
®
.
4. Moisten a clean, lint-free cloth with denatured alcohol and clean lamp(s).
5. When
STER-L-RAY
®
lamp(s) have been removed, cleaned and safely set aside, remove any loose dust or dirt from the
interior of the
Nutripure
®
with a soft dry cloth, or vacuum.
6. To remove any dust, dirt, fingerprints, smears, etc, from the outer surfaces of the conditioner, moisten a clean, lint-free cloth
with denatured alcohol and wipe down the conditioner. Buff out any streaks, left from the cleaning, using a soft dry cloth.
7. Install
STER-L-RAY
®
lamp(s), insert lamp pin into spring-loaded lampholder. Carefully press
STER-L-RAY
®
lamp against
the spring-loaded lampholder. Align lamp pin with stationary lampholder. Slowly ease the pressure until
STER-L-RAY
®
lamp is fully seated in lampholders.
8. Verify that
STER-L-RAY
®
lamp(s) is properly seated.
9. Repeat until all
STER-L-RAY
®
lamps have been cleaned and properly reinstalled.
10. Install Access Cover. Restore power to the
Nutripure
®
.
Avoid exposure to direct or reflected germicidal ultraviolet rays. Germicidal
ultraviolet rays are harmful to the eyes and skin.
11. Wear personal protection equipment, then verify operation by viewing
STER-L-RAY
®
lamps(s). If the fixture is not
operating properly, go to
“Troubleshooting”
section.
Summary of Contents for Nutripure 2B-SC
Page 13: ...13 MAINTENANCE NOTES ...
Page 14: ...14 MAINTENANCE NOTES ...
Page 15: ...15 MAINTENANCE NOTES ...