3.2
Flow Tee Mounting
Convertible sensors may be used in a 1” flow tee as shown in Figure 7. The flow tee is a
modified pipe fitting that accommodates the pipe thread on the front of the sensor. Sample must
flow directly against the face of the sensor as shown. The sensor may be mounted horizontally
provided that the outlet flow is pointed up to avoid “air locking” in the tee. Note that standard 1”
tee fittings will not work without modification due to clearance problems in most molded tees.
Figure 7 - Flow Through Tee Mount