10
O&M Manual
Rev-F (8/17)
Part 2 – Analyzer Mounting
2.1
General
All Q46 Series instruments provide for mounting flexibility. A bracket is included that allows
mounting to walls or pipes. In all cases, choose a location that is readily accessible for
calibrations. Also consider that it may be necessary to utilize a location where solutions can be
used during the calibration process. To take full advantage of the high contrast display, mount
the instrument in a location where the display can be viewed from various angles and long
distances.
Locate the instrument in close proximity to the point of sensor installation - this will allow easy
access during calibration. The standard cable length of the pH sensor is 15 feet. For sensor
cables longer than 30 feet, use the optional junction box (07-0100) and sensor interconnect cable
(31-0057).
Figure 1 – Q46 Enclosure Dimensions