ATI
Q46H/79PR Chlorine Manual
P
art 4 – Electrical Installation
20
O&M Manual
Rev-E (5/17)
4.2
Power Connection
Verify the AC power supply requirement before installing. Also verify that power is fully
disconnected before attempting to wire.
AC powered Q46 systems are supplied with 5 cable gland fittings.
Connect HOT, NEUTRAL, and GROUND to the matching designations on terminal strip
TB7
.
The two analog outputs for the standard system are present at terminal TB1. The loop-load
limitation in this configuration is 500 Ohms maximum for each output. Also note that these two
outputs are completely isolated from each other to insure that ground loops do not result from the
connection of both outputs to the same device such as a PLC or DCS.
A ribbon cable connects the power supply assembly with the microprocessor assembly located in
the front section of the enclosure. This cable can be removed during installation to facilitate
wiring if desired. It is best to unplug only one end. The ribbon cable has a marking stripe on one
edge that is used to indicate proper orientation. The indicator stripe should be on the bottom
edge of the cable when installed.
WARNING
Disconnect line power voltage BEFORE connecting line power
wires to Terminal TB7 of the power supply. The power supply
accepts only standard three-wire single phase power. AC power
supplies are configured for 90-260 VAC operation at the factory
at time of order, and the power supply is labeled as such. Do
NOT connect voltages other than the labeled requirement to the
input.