ATI
Q46H/63 Combined Chlorine System
Part 4 – Electrical Installation
26
O&M Manual
Rev-K (11/18)
4.4
Sensor Wiring
The sensor cable can be quickly connected to the Q46 terminal strip by matching the wire colors
on the cable to the color designations on the label in the monitor. A junction box is also available
to provide a break point for long sensor cable runs. Route signal cable away from AC power
lines, adjustable frequency drives, motors, or other noisy electrical signal lines. Do not run
sensor or signal cables in conduit that contains AC power lines or motor leads
Note: If sensor is experiencing Low-Slope or Low-Output conditions, due To poor Earth
Ground Connections, move the Shield connection from P/S board to alternate location on
lid, where indicated with an “S”.
Figure 18 - Direct Sensor Connection
Note
: The
BLUE
wire is
NOT
used when connecting a Flow or
Submersible style Probe to the
analyzer
Note:
pH input signal may be supplied
from a separate isolated voltage input
of 0.3-1.5 VDC
(75 Ohm resistance across a 4-20 output
)