background image

ATI  

Q46H/63 Combined Chlorine System 

Part 1 - Introduction 

 

O&M Manual 
Rev-K (11/18) 

 
1.3 

Features 

 

 

Standard  Q46H/63  analyzers  have  fully  isolated  inputs  and  outputs.    Analog  outputs  are 
additionally completely isolated from each other. 
 

 

Available  in  either  90-260  VAC  or  12-24  VDC  power  supply  systems.    All  features  remain  the 
same in both variations.   

 

 

High  accuracy,  high  sensitivity  system,  measures  from  20  ppb  to  200.0  ppm  through  4  internal 
automatic ranges.  User ranges of 200.0 ppb, 2.000 ppm, 20.00 ppm, or 200.0 ppm.  

 

 

Output Hold, Output Simulate, Output Alarm, and Output Delay Functions.  All forced changes in 
output condition include bumpless transfer to provide gradual return to on-line signal levels and to 
avoid system control shocks on both analog outputs. 

 

 

Three  6  amp  SPDT  relay  outputs  and  two  analog  4-20  mA  outputs  are  standard.    Software 
settings  for  relay  control  include  setpoint,  deadband,  phase,  delay,  and  failsafe.    An  optional  3-
relay card, for 0-30 VDC signals, is available to bring the total to 6 relays.   

 

 

Selectable  PID  controller  on  main  analog  output.    PID  controller  can  operate  with  instrument 
configured  as loop-power  analyzer or as one  of the two outputs on the  AC  powered instrument.  
PID  includes  manual  operation  feature,  and  diagnostic  “stuck-controller”  timer  feature  for  alarm 
output of control problems. 
 

 

Digital  communication  options  for  Profibus-DP,  Modbus  RTU.  More  versions  pending.  See  Q46 
Digital Communications Manual for Specifications. 

 

 

Optional pH sensor feature enables active pH measurement with the Q46H/63 monitor.  The pH 
reading can also be sent to one of the analog outputs for complete pH and chlorine monitoring. 
 

 

Two analog outputs may be configured to track chlorine and temperature, chlorine and chlorine, 
or chlorine and pH.  Both analog outputs can be individually programmed to fail to specific values.  
An  optional  third  analog  output  is  available,  providing  separate  outputs  for  chlorine,  pH,  and 
temperature.   

 

 

Selectable  Output  Fail  Alarm  feature  allows  system  diagnostic  failures  to  be  sent  to  external 
monitoring systems. 

 

 

Large, high contrast, custom LCD display with LED back light provides excellent readability in any 
light conditions.  The secondary line of display utilizes 5x7 dot matrix characters for clear 
message display.  Two of four measured parameters may be on the display simultaneously.   

 

 

Diagnostic messages provide a clear description of any problem with no confusing error codes to 
look up.  Messages are also included for diagnosing calibration problems. 

 

 

Quick and easy one-point calibration method and sensor zero-cal.  To provide

 

high accuracy, all 

calibration methods include stability monitors that check temperature and main parameter stability 
before accepting data. 

 

 

High accuracy three-wire Pt100 temperature input. Temperature element can be user calibrated. 

 
 

Summary of Contents for Q46H/62

Page 1: ...nology Inc ATI UK Limited 6 Iron Bridge Drive Unit 1 2 Gatehead Business Park Collegeville PA 19426 Delph New Road Delph Phone 800 959 0299 Saddleworth OL3 5DE 610 917 0991 Phone 44 0 1457 873 318 Fax...

Page 2: ...TWARE NAVIGATION 36 6 22 MEASURE MENU MEASURE 38 6 23 CALIBRATION MENU CAL 39 6 24 CONFIGURATION MENU CONFIG 40 6 25 CONTROL MENU CONTROL 45 6 26 DIAGNOSTICS MENU DIAG 49 PART 7 CALIBRATION 52 7 1 CHL...

Page 3: ...SEMBLY 19 FIGURE 13 LINE POWER CONNECTION 22 FIGURE 14 OPTIONAL RELAY BOARD WIRING 23 FIGURE 15 OPTIONAL RELAY BOARD WIRING 24 FIGURE 16 OPTIONAL ANALOG OUTPUT WIRING 24 FIGURE 17 SENSOR CABLE PREPARA...

Page 4: ...ed chlorine monitor is a polarographic membrane sensor which measures chlorine directly Water simply flows past the sensor and directly to drain with the flow rate and pressure across the sensor contr...

Page 5: ...07 0096 pH sensor For connection of the sensor to the electronics a 25 foot cable is supplied An additional 100 feet of interconnect cable may be added using 07 0100 junction box All required spare p...

Page 6: ...Figure 2 below shows the same standard flowcell assembly and chlorine sensor along with the conventional type pH sensor A special adapter is required to hold the pH sensor in its proper location in t...

Page 7: ...nd the use of our 03 0372 flow control assembly when using sealed flowcells This assembly consists of an in line filter and a fixed flow regulator which will maintain a constant 400 cc min flow rate t...

Page 8: ...control problems Digital communication options for Profibus DP Modbus RTU More versions pending See Q46 Digital Communications Manual for Specifications Optional pH sensor feature enables active pH me...

Page 9: ...to 200 0 ppm Display 0 75 19 1 mm high 4 digit main display with sign 12 digit secondary display 0 3 7 6 mm 5x7 dot matrix Integral LED back light for visibility in the dark Keypad 4 key membrane typ...

Page 10: ...124 mm x 4 9 124 mm x 5 5 139 mm Mounting Options Wall or pipe mount bracket standard Bracket suitable for either 1 5 or 2 I D U Bolts for pipe mounting Panel mount adapter optional Conduit Openings...

Page 11: ...d range Non linearity 0 1 of selected range Warm up Time 3 seconds to rated performance electronics only Supply Voltage Effects 0 05 span Instrument Response Time 60 seconds to 90 of step input at low...

Page 12: ...ocess To take full advantage of the high contrast display mount the instrument in a location where the display can be viewed from various angles and long distances Locate the instrument in close proxi...

Page 13: ...s then attached to the wall Figure 6 using the four outer mounting holes in the bracket These holes are slotted to accommodate two sizes of u bolt that may be used to pipe mount the unit Figure 7 Slot...

Page 14: ...ATI Q46H 63 Combined Chlorine System Part 2 Analyzer Mounting 14 O M Manual Rev K 11 18 Figure 6 Wall Mounting Diagram Figure 7 Pipe Mounting Diagram...

Page 15: ...the rear of the enclosure 4 screws for attachment of this bracket and a sealing gasket to insure that the panel mounted monitor provides a water tight seal when mounted to a panel The sealing gasket m...

Page 16: ...low concentrations below 0 5 PPM this method should always be used Mechanical installation of the flow cell requires that it be mounted to a wall or other convenient flat surface Alternatively the mo...

Page 17: ...well controlled can use a simpler sealed flowcell Using this flowcell requires that flow be controlled externally to about 400 cc min Variable flow rate or variable pressure will cause unstable readi...

Page 18: ...ATI Q45H 63 Chlorine Manual Part 3 Sensor Flowcell Mounting 18 O M Manual Rev K 11 18 Figure 11 Sealed pH Flowcell Details...

Page 19: ...ns for a submersible chlorine sensor should first be discussed with ATI A trial of such installations may be necessary Submersible sensors are mounted to a 1 pipe using a standard 1 PVC thread by thre...

Page 20: ...oise from coupling with the instrumentation signals 3 This analyzer must be installed by specifically trained personnel in accordance with relevant local codes and instructions contained in this opera...

Page 21: ...ction of both outputs to the same device such as a PLC or DCS A ribbon cable connects the power supply assembly with the microprocessor assembly located in the front section of the enclosure This cabl...

Page 22: ...Part 4 Electrical Installation 22 O M Manual Rev K 11 18 Figure 13 Line Power Connection The power strip TB7 allows up to 12 AWG wire A wire gauge of 16 AWG is recommended to allow for an easy pass th...

Page 23: ...the contacts For applications that require the same switched operating voltage as the Q46 115 or 230 V power may be jumpered from the power input terminals at TB7 Relay wiring is connected at TB4 TB5...

Page 24: ...to connect to the optional 3 relay card Figure 15 OR the optional third analog output Out 3 Figure 16 The Q46 can be configured for only one of these optional features and the hardware for either opti...

Page 25: ...removed easily from the flowcell Cord grips used for sealing the cable should be snugly tightened after electrical connections have been made to prevent moisture incursion When stripping cables leave...

Page 26: ...frequency drives motors or other noisy electrical signal lines Do not run sensor or signal cables in conduit that contains AC power lines or motor leads Note If sensor is experiencing Low Slope or Low...

Page 27: ...monitor max 100 feet require the use of a junction box The junction box is shown in and is supplied with two cable glands on the bottom of the enclosure Note If sensor is experiencing Low Slope or Lo...

Page 28: ...ignal of 0 3 1 5 VDC A standard pH sensor without preamp 63 0101 is also available but the maximum cable length for this sensor is 25 ft 7 5 m When this sensor is used with the constant head flowcell...

Page 29: ...ned Chlorine System Part 4 Electrical Installation 29 O M Manual Rev K 11 18 QUANTUM Figure 20 Junction Box Wiring w Optional pH Sensor Note The BLUE wire is NOT used when connecting a Flow Style Prob...

Page 30: ...rocedure should be done by a qualified technician and it should only be done when the system is ready for operation Until then it is best to leave the sensor in the condition in which it is received F...

Page 31: ...talled with the proper side facing the sample Examine the membrane carefully to identify the shiny side The membrane has a dull side and a shiny side The shiny side must face out or be in contact with...

Page 32: ...with the Q46H system one battery powered and one without an internal preamp The pH sensors may be used to provide for continuous monitoring of both combined chlorine and pH Both sensors are shipped wi...

Page 33: ...d from this user interface no internal jumpers pots etc MENU ICONS UNITS 12 CHARACTER SECONDARY DISPLAY MEMBRANE KEYPAD MENU ESC ENTER A B DIAG FAIL HOLD CAL CONF MENU ICONS UNITS 12 CHARACTER SECONDA...

Page 34: ...12 Display The large custom display provides clear information for general measurement use and user configuration There are three main areas of the display the main parameter display the secondary mes...

Page 35: ...In this case the output is locked to the last input value measured when the HOLD function was entered HOLD values are retained even if the unit power is cycled FAIL The FAIL icon indicates that the s...

Page 36: ...en completed pressing ENTER again stores the variable and stops the flashing Pressing ESC aborts the change and also exits user entry mode The starting default screen is always the Measure Menu The UP...

Page 37: ...elay C Mode 5 Relay E Mode 5 Relay D Mode 5 Relay F Mode 7 Com Mode 7 Com Address 4 Set 4mA 3 4 Set 20mA 3 6 Setpnt A or A HI A LO 6 Hyst A or A HI A LO 6 Delay A or A HI A LO Phase A Setpnt B Hyst B...

Page 38: ...mA PID Status screen if enabled Shows the present controller output level on left and actual analyzer current on the right The controller can be placed in manual while viewing this screen by pressing...

Page 39: ...solution or to set the sensor zero point See Part 7 Calibration for more details Cal pH The pH calibration function allows the user to if enabled adjust the analyzer offset and span to match the samp...

Page 40: ...dify value range is 0 1 to 9 9 minutes Press ENTER to store the new value Contrast This function sets the contrast level for the display The custom display is designed with a wide temperature range Su...

Page 41: ...for temperature default chlorine or pH Press ENTER to initiate user entry mode and the entire value will flash Use the UP arrow key to modify the desired value selections include 1 C F for temperatur...

Page 42: ...ay B then operates identically to Relay A with settings for setpoint hysteresis delay and phasing appearing in the CONFIG menu automatically See Figure 25 for details Temp Units This function sets the...

Page 43: ...automatically identify which buffer is being used and compensate for the value based on the built in tables The order in which the buffers are used during calibration is unimportant since the system a...

Page 44: ...4 24 95 7 01 95 9 77 4 00 pH 7 00 pH 9 18 pH C pH C pH C pH 0 4 00 0 7 10 0 9 46 10 3 99 10 7 06 10 9 33 20 4 00 20 7 02 20 9 23 30 4 01 30 6 99 30 9 14 40 4 03 40 6 97 40 9 07 50 4 05 50 6 98 50 9 01...

Page 45: ...the range specified in Set Range and the 0 and the 100 point must be separated by at least 1 of this range Use the LEFT arrow key to select the first digit to be modified Then use the UP and LEFT arr...

Page 46: ...may be set to track temperature default pH or chlorine The values stored for the 4 mA point may be higher or lower than the value stored for the 20 mA point The entry value is limited to a value betw...

Page 47: ...between 0 and 999 seconds Press ENTER to initiate user entry mode and the value will flash Use arrow keys to modify value range is 0 to 999 seconds Press ENTER to store the new value A Phasing This fu...

Page 48: ...entically to Relay A Relay B settings appear in the CONFIG menu list automatically If Relay C Mode is set to CON see Relay C Mode then Relay C will function identically to Relay A Relay C settings app...

Page 49: ...the new value Note the Set Hold function holds BOTH current levels as well as ALL relay settings The Set Hold function can also hold at an output value specified by the user To customize the hold valu...

Page 50: ...bleshooting Escaping this screen returns the unit to normal operation Press ENTER to initiate the user entry mode and the right most digit of the value will flash Use arrow keys to modify desired valu...

Page 51: ...imit is exceeded but also when power is lost to the controller When failsafe is selected to be ON the normally open contacts of the relay will be closed during normal operation In an attempt to make t...

Page 52: ...by plugging the cable plug into the receptacle on the top of the sensor 2 Place about an inch of water in a small beaker or other convenient container and immerse the tip of the sensor The water used...

Page 53: ...into place If insertion becomes difficult use a small amount of silicon grease to lubricate the o rings If the low volume flow cell is used screw the sensor into the flow cell until the membrane cap...

Page 54: ...may be calibrated on line or the sensor can be removed from the process and placed into a known solution temperature reference In any case it is critical that the sensor be allowed to reach temperatur...

Page 55: ...e a problem with the temperature element In the event of calibration failure it is recommended to attempt the calibration again immediately 7 3 pH Calibration The pH calibration menus will not be seen...

Page 56: ...he effect It is not possible to eliminate this effect however it usually is constant or very slow changing A full calibration of the pH sensor consists of an initial 2 point calibration in buffers Thi...

Page 57: ...hly clean the pH sensor and place it into the pH calibration buffer stirring several times before letting it rest in the beaker Allow sensor to sit in solution long enough to achieve temperature equil...

Page 58: ...to keep the process at the setpoint To affect this control the controller must be used with properly selected control elements valves proper chemicals etc that enable the controller to add or subtract...

Page 59: ...of the controller Like proportional gain increasing integral gain results in the control action happening quicker Set the I term to a number between 3 5 to start 1 2 more than P Like proportional gai...

Page 60: ...ss and hold the ENTER key for about 5 seconds You will see a small m show up between the value and the mA value This indicates you are now in manual mode Once in manual you may increase the PID output...

Page 61: ...at the extreme ranges of the measuring scale In addition if a chemical process is not symmetrical that means it responds differentially to the addition and subtraction of chemical It is important in...

Page 62: ...ifficult to clean effectively Immersing the tip of the sensor in 1N nitric acid solution will sometimes remove deposits that cause low sensitivity but this is not always the case The recommended pract...

Page 63: ...the tip and the silver coil Do not allow the connector at the back of the sensor to be submerged Allow the electrodes to soak in distilled water for 30 minutes 4 Put a new membrane and fresh electrol...

Page 64: ...ing with any acid regardless of concentration 9 6 Replacing the Saltbridge and Reference Buffer Solution 1 Hold the sensor with the process electrode pointing up Place a cloth or towel around the salt...

Page 65: ...r without interfering with operations The flow cell may be cleaned by wiping or by washing with detergents or dilute acids Do not try to clean with solvents as the acrylic may craze or crack Change th...

Page 66: ...leshooting it is many times helpful to disconnect all these outputs and place wire shorts across the terminals in the instrument to isolate the system and evaluate any problems which may be coming dow...

Page 67: ...e in the water is combined chlorine and not free chlorine 10 3 Analyzer Tests 1 Disconnect power and completely disconnect all output wiring coming from the analyzer Remove sensor wiring relay wiring...

Page 68: ...configured Check manual for limits of the function to be configured Locked Analyzer security setting is locked Enter security code to allow modifications to settings Unlocked Analyzer security has ju...

Page 69: ...sor tests as described in section 10 5 Replace sensor if still failure TC Cal Fail Failure of temperature calibration FAIL icon will not extinguish until successful calibration has been performed or 3...

Page 70: ...d an open circuit value of infinite resistance There must be no measurable resistance at all between these wires Any resistance at all indicates either water in the cable connector or the breakdown in...

Page 71: ...ng replace the saltbridge and reference cell buffer pH 7 as outlined in Section 9 If the sensor still cannot be calibrated perform the following test A multimeter pH 7 buffer and pH 4 buffer will be n...

Page 72: ...erence Solution for pH sensor 07 0096 05 0066 Salt bridge for pH sensor 07 0096 00 0043 Constant Head Flowcell assembly with mounting plate 42 0014 Flowcell o ring each 07 0100 Junction box 31 0038 In...

Page 73: ...e alteration improper storage or extended exposure to excessive gas concentrations Should inspection indicate that sensors have failed due to any of the above the warranty shall not apply The Manufact...

Page 74: ...DUCTS NH3 Ammonia CO Carbon Monoxide H2 Hydrogen NO Nitric Oxide O2 Oxygen CO Cl2 Phosgene Br2 Bromine Cl2 Chlorine ClO2 Chlorine Dioxide F2 Fluorine I2 Iodine HX Acid Gases C2H4O Ethylene Oxide C2H6O...

Reviews: