Q46/84 Hydrogen Peroxide Monitor
Part 2– Analyzer Mounting
12
O&M Manual
Rev-H (6/17)
2.3
Panel Mounting
Panel mounting of an AC powered monitor uses the panel mounting flange molded into the rear
section of the enclosure. Figure 6 provides dimensions for the panel cutout required for
mounting. Loop-powered instruments are normally not panel mounted, but may be panel
mounted if ordered specifically for this purpose. Panel mounted 2-wire systems require the use
of the “deep back” enclosure normally used for AC powered systems.
The panel mounting bracket kit must be ordered separately (part # 05-0068). This kit contains a
metal retainer bracket that attaches to the rear of the enclosure, 4 screws for attachment of this
bracket, and a sealing gasket to insure that the monitor provides a water tight seal when mounted
to a panel.
The sealing gasket must first be attached to the enclosure. The gasket contains an adhesive on
one side to remains in place on the enclosure. Remove the protective paper from the adhesive
side of the gasket and slide the gasket over the back of the enclosure so that the adhesive side
lines up with the back of the enclosure flange. Once in place, proceed to mount the monitor in
the panel.
Figure 6 - Panel Mount and Cut-Out