QSG_F600HP_425.00_EN_01_2021-01-25
30
/
34
Additional functions
Some functions are available depending on software version
.
Label
Function
Description
24V OUTPUTS
Auxiliaries output 24 V
Available outputs for external automatism
ABSOLUTE
Absolute
Display the absolute value of the results
ATF
ATF time
Absorb the important leak variations at the
defined time available with flow units only
ATR0
/
ATR1
/
ATR2
/
ATR3
ATR 0 - 3
Specific filters on leak measurement
BUZZER
Buzzer
Buzzer activation configuration
CODE READER
Bar code reader
Bar code configuration
CUT OFF
Cut off
All the measurements that are lower than the
configured rate have the value 0
DISP. OPT.
Display option
Display of an additional information on a second
line
DISPLAY MODE
Display Mode
Leak measurement resolution
DUMP OFF
Dump off
Avoids dumping (when activated in Configuration /
Pneumatic)
END OF CYCLE
End of cycle
Several automatism case depending on fail part
management
FILL MODE
Fill types
Special filling methods
FILTER
Filtering
Stabilize the measurement values
FLOW LEVEL
Flow level
Add a minimum fail parameter
NAME
Name
Program customization
NO NEGATIVE
No Negative
Replace negative value per 0
OFFSET
Leak offset
Leak offset value
PEAK HOLD
Peak hold
Give as result, the highest flow during the test
time
PR:SEQUENCE
Sequencing
Allowed program automatic sequencing
PRESS.CORR.
Pressure correction
Calculates leak at a defined pressure value (flow units)
PRESSURE DROP
Pressure drop
Pressure drop mode function in the Desensitized
mode
REWORK LIMIT
Rework limits
Additional levels for specific reworkable parts
SIGN
Sign
Return opposite result
STAMPING
Stamp
Pneumatic or electric output to identify the part
STD CONDITIONS
Standard conditions
Standard conditions correction with parameters (flow
units)
TEST TIME*100
Longer test time
Allowed longer test time (1s = 100s)
UNITS
Units
Access to International System or American or
Custom Units
VALVE CODES
Valve codes
Available outputs for external automatism