PARAMETERS menu
Use this menu to configure the measurement cycle associated to each test program.
Default parameters of the type tests
Label
Parameter
Description
COUPL. A
Coupling time
Waiting time before filling the test part
FILL TIME
Fill time
Time to pressurise the part under test
STAB TIME
Stabilization time
Time to stabilise the pressure on the test and
reference parts
TEST TIME
Test time
Time for leak measurement
DUMP TIME
Dump time
Time to vent the part to atmosphere
Press. UNIT
Pressure units
Pressure unit (Pa, kPa, MPa, bar, mbar, PSI,
mmHg, inHg, Pts)
Max FILL
Maximum fill pressure
Maximum level of the fill pressure
Min FILL
Minimum fill pressure
Minimum level of the fill pressure
LeakUnit
Reject unit
Measurement units (kPa, kPa/s, MPa, bar, mbar,
PSI, cm
3
/s, cm
3
/min, cm
3
/h, mm
3
/s, ml/s, ml/min,
ml/h, l/min, H
2
0 µG/min, Pts)
Test FAIL
Test fail
Upper leak rate limit for the test part. Above this
limit, the part is considered as defective.
FUNCTIONS
Functions
Access to additional functions
Some parameters are displayed when specific functions are activated
.
Label
Parameter
Description
INTER-CYCLE
Inter cycle time
Time between two automatically chained
programs (Sequence function)
OFFSET
Leak offset
Leak offset value
REJECT CALC.
Reject calculation
Define raw unit to calculate flow unit (Flow unit)
Set FILL
Set fill
Fill pressure instruction (Fill function or
electronic pressure regulator)
VOLUME
Test volume
Complete volume of the test part (Flow unit)
Volume UNIT
Volume unit
Volume unit of the test part (Flow unit)
Max Fail
Maximum fail
Test FAIL is replaced by this one when the function
Flow Level = YES
Min Fail
Minimum fail
Appears when the function Flow Level = YES
Test REWORK
Test rework
Test for specific reworkable parts