background image

 

 

 
 
 

ATEQ F600HP 

Quick Start Guide

Summary of Contents for F600HP

Page 1: ...ATEQ F600HP Quick Start Guide...

Page 2: ...anel 8 Connectors on the back panel with all options 9 Power supply connectors 11 Digital links 12 Analog outputs option 15 Digital inputs outputs 16 Pneumatic connectors 19 Pneumatics configuration 2...

Page 3: ...ateq com cn T 86 21 6763 9508 F 86 21 6763 9528 ATEQ SYSTEMS ANALYSIS TAIWAN CO LTD NO 3 LAN 223 San Jia Dong Street 40642 TAICHUNG TAIWAN ateqtaiwan ateq com tw ateq com tw T 886 4 2437 5278 F 886 4...

Page 4: ...ad the user manual before using the device All electrical connections to the device must be equipped with safety systems fuses circuit breakers etc adapted to the needs and in accordance with the appl...

Page 5: ...itions impurities must be prevented from being drawn into its internal components For this purpose we strongly recommend that a suitable airtight filter is installed between the part under test and th...

Page 6: ...F600HP_425 00_EN_01_2021 01 25 6 34 Preamble ATEQ F600HP AN UNIVERSAL LEAK TESTER ATEQ F600HP is a leak detector that tests the air tightness of parts ATEQ F600HP can memorize 128 different test progr...

Page 7: ...The measurement is performed by the pressure sensor 2 1 Device 2 Pressure sensor 3 Part under test Desensitized Mode there is no leak fullscale The limit is the test pressure PRINCIPLE OF A CYCLE The...

Page 8: ...21 01 25 8 34 Your ATEQ F600HP FRONT PANEL The user interface is located on the front panel 1 Display 2 Cycle keys 3 Navigation keys 4 USB connectors 5 Mechanical Pressure Regulator For more informati...

Page 9: ...QSG_F600HP_425 00_EN_01_2021 01 25 9 34 CONNECTORS ON THE BACK PANEL WITH ALL OPTIONS...

Page 10: ...gh pressure connector 6 High pressure air input connector to the metallic filter option 7 Pressure range 8 Part number Serial number 9 Air supply energy information 10 Pilot pressure input 11 Ground 1...

Page 11: ...connector J7 The device can be connected to a 24 V DC 2 A power supply through a M12 4 pins type connector 24 V DC on the relay board connector J11 option The device can also be connected to a 24 V D...

Page 12: ...SB connectors on front face USB connectors can be used for connecting miscellaneous compatible USB devices The USB connectors are located under the rubber cover 1 see figure 1 Rubber cover 2 USB conne...

Page 13: ...RS232 SubD 9 pins female connector Profibus option Profibus SubD 9 pins female connector Pin number Signal 1 Not used 2 RXD data input 3 TXD data input 4 Not used 5 Ground 6 Not used 7 RTS request to...

Page 14: ...ins female connector J6 Devicenet output For connection to other ATEQ devices Profinet connectors J5 J6 option M12 D coded type connector 4 pins female connector J5 J6 Pin number Signal 1 Drain 2 V 3...

Page 15: ...these network protocols is available Ethernet IP Profinet Ethercat ANALOG OUTPUTS OPTION M12 type connector 8 pins Female connector J1 Pin number Description 1 Ground pressure 2 0 10 V DC Pressure 3...

Page 16: ...60 V AC DC max 200 mA max The device can be energized through the J11 connector of the relay board except if internal supply option 0 V to the pin 16 24 V DC to the pin 2 or 4 Pin number Inputs output...

Page 17: ...of the program number Pin number Inputs outputs Description 1 I8 Input 8 Program selection from 33 to 64 programmable input 2 I9 Input 9 Program selection from 65 to 128 programmable input Program num...

Page 18: ...4 V DC Common outputs 1 2 3 2 Output 1 Open collector 3 Output 2 Open collector 4 Output 3 Open collector 5 24 V DC Common outputs 4 5 6 6 Output 4 Open collector 7 Output 5 Open collector 8 Output 6...

Page 19: ...anel of the device Pneumatic supply The pneumatic supply has to meet specific requirements recommended by ATEQ Refer to Good practices and safety instructions section A specific filter may be necessar...

Page 20: ...st output with a silencer Air supply input 600kPa Air supply input for the internal pilot pressure 600 kPa 87 PSI 15 Instant fitting 6 mm diameter Test air supply input Test high pressure input Filter...

Page 21: ...gulator air supply to 1 Connection of the test high pressure input Filter is optional if clean air is supplied or air supply input for the internal high pressure regulator option Air supply to 2 Conne...

Page 22: ...the front panel 1 Display 2 Cycle keys 3 Navigation keys KEYS Cycle keys The cycle keys are used to start and to stop a measurement cycle Key Name Function Start On the Program screen starts a measure...

Page 23: ...reen escapes without modifying parameters Smart key Smart key is a programmable key that provides direct access to a function selected by the user Key Name Function Smart key Starts a measurement cycl...

Page 24: ...USB connection 11 Current program 12 Vertical line test result The MAIN MENU screen The MAIN MENU screen gives access to different sections for managing the device and the test parameters Access from...

Page 25: ...am screen is displayed with last program used on screen PREPARING A PROGRAM Use this procedure to configure a new test program On the MAIN MENU screen ACCESSING THE PARAMETERS 1 Select PARAMETERS usin...

Page 26: ...icon is displayed at the bottom of the screen you must insert the USB unlocking device or enter a password before modifying a parameter On the PARAMETERS screen of the program see Preparing a program...

Page 27: ...ogram displayed on the Program screen STARTING A MEASUREMENT CYCLE 1 Press Start The cycle phases of the program are successively displayed FILL STABILISATION TEST DUMP At the end of the cycle the res...

Page 28: ...ent Pressurize the part and allow to adjust pressure Levels Infinite Fill Infinite fill Pressurize the part with an infinite fill time Piezo auto zero Piezo auto zero Auto zero cycle on the piezo sens...

Page 29: ...MPa bar mbar PSI cm3 s cm3 min cm3 h mm3 s ml s ml min ml h l min H20 G min Pts Test FAIL Test fail Upper leak rate limit for the test part Above this limit the part is considered as defective FUNCTIO...

Page 30: ...automatism case depending on fail part management FILL MODE Fill types Special filling methods FILTER Filtering Stabilize the measurement values FLOW LEVEL Flow level Add a minimum fail parameter NAME...

Page 31: ...program that becomes 0 second AUTOMATISM Automatism Configuration of the different communications between the device and its environment RS232 Configuration of the communication type on the RS232 port...

Page 32: ...lows to display more special cycles to carry out device internal tests RESET PARA Parameters reset Reset to factory configuration RESULTS menu In this section manage measurements results USB the resul...

Page 33: ...his section describes save and restore parameters on an external USB device Label Description Save parameters Save parameters on an external USB memory device for restoring later Restore parameters Re...

Page 34: ...91 x 9 84 x 10 63 Overall dimensions 150 x 250 x 360 mm 5 91 x 9 84 x 14 17 Mass About 8 kg 17 6 lbs Electrical power supply 100 240 V AC 50 W 50 60 Hz 24 V DC 2 A Overvoltage category II Protection D...