.394.00_EN_01 / 2019-02-25
21 / 42
Test and reference outputs
The outputs enables parts to be connected (test and reference)
1
Test connector
2
Reference connector
3
Not used
4
Exhaust output
5
Pressurization output
Metallic fitting available for test (1
) and
reference (2
) connectors:
—
2.7/4 mm
—
3/5 mm
—
4/6 mm
—
6/8 mm
Quick connector (option)
Use this function to check the calibration.
As this connector is part of the measurement circuit, all its connections must be air tight.
Calibration check by volume variation connector (option) (V)
External volume (closed tube) connection.
Summary of Contents for F28+
Page 1: ...ATEQ F28 Quick Start Guide...