.394.00_EN_01 / 2019-02-25
16 / 42
DIGITAL INPUTS/OUTPUTS
The 24V DC power supply for the digital inputs can be provided by 2 means:
—
The internal power supply of the device (0.3A max)
—
An external power supply provided by the customer.
Inputs default mode is PNP. NPN mode is available on request.
Relay board connector (J8) (option)
Characteristics
—
Inputs
•
Activation: + 24 V DC.
—
Outputs
•
Dry contacts
•
60 V AC / DC max - 200 mA max.
Pin number
Inputs / outputs
Description
1
Input 1
RESET
2
+ 24 V DC
Common
3
Input 2
START
4
+ 24 V DC
Common
5
Input 3
Program selection
6
Input 4
Program selection
7
Input 5
Program selection
8
Input 6
Program selection
9
Input 7
Program selection (programmable input)
10
Output
Common floating output
11
Output
Pass part
12
Output
Test fail part
13
Output
Reference fail part
Summary of Contents for F28+