QSG_ERD620_421.00_EN_01 / 2020-09-09
38
/
48
The Measurement cycle screen
The
Measurement cycle
screen displays the different values of the current test (or last one).
1
Test pressure measurement
2
Program name
3
Test result or step phase
4
Flow reject value
5
Vertical line test result*
6
Step flow measurement (when
the function Step flow is
selected)
7
Remaining time of the current phase
or ready status
8
Down pressure event (when the
function Pressure Down is
selected)
9
Event pressure unit
10
Up pressure event (when the
function Pressure Up is selected)
11
Current program
* Indicates the result: both green = Pass, red on the left = Fail Up, red on the right = Fail
Down, both red bottom = step flow low, both red up = step flow high, both red = pressure
alarm
The MAIN MENU screen
The
MAIN MENU
screen gives access to different sections for managing the device and the
test parameters.
Access: from the
Program
screen, press
.
Option
Description
SPE CYCLE
Specific procedures necessary to ensure the proper operation of
measurement cycles (for example, adjustment of a pressure regulator).
PARAMETERS
Parameters of the test programs.
CONFIGURATION
General configuration of the device.
SERVICE
Maintenance of the device.
RESULTS
Test results, backup and display options.
USB
USB connection functions (backup, restore).