QSG_ERD620_421.00_EN_01 / 2020-09-09
44
/
48
Additional functions
Label
Function
Description
24V OUTPUTS
Auxiliaries output 24 V
Available outputs for external automatism
ANALOG OUTPUT
Analog output
Duplicate the flow and pressure measurements on
the analogues outputs
(Window mode is only for
Direct or Indirect Flow Test and do not use with
ERD)
AUTO CONNECT
Automatic connector
Function to manage automatic jigs
BUZZER
Buzzer
Buzzer activation configuration
CODE READER
Code reader
Bar code reader configuration
DISPLAY MODE
Display Mode
Flow measurement resolution
DN CONTACT
Down contact
Contact type parameter
between the inferior and
superior pressure limits during the drop, open or
close. It is the same pressure levels as for flow
event.
END OF CYCLE
End of cycle
Several automatism case depending on fail part
management
(Only for Direct or Indirect Flow Test
and do not use with ERD)
FILTER
Filtering
Stabilize the measurement values
MEAS. DELAY
Measurement delay
During this time reject levels are not controlled
(perturbations at starting ramp)
MINI-VALVE
Mini valve
Access to highest time resolution (fast test) and
auto zero time
NAME
Name
Program customization
OFFSET
Offset
3 Flow offset: Offset Up, Offset Step, Offset
Down
PR:SEQUENCE
Sequencing
Allowed program automatic sequencing
PRESS. AZ
Pressure auto zero
Define if Pressure auto zero is done before, after
the cycle or on demand (not done automatically)
PRESS. CORR.
Pressure correction
Calculate flow measurement at a specific
pressure value, whatever the real test pressure is
PRESS. DOWN
Pressure down
Inferior and superior pressure limits and maximal
flow value during the dropping phase (flow event)
PRESS. UP
Pressure up
Inferior and superior pressure limits, minimal flow
value during the rising phase (flow event)
SHORT CYCLE
Cycle shortened
When fail is detected, the test is stopped and
does the dropping ramp
SIGN
Sign
The flow sign is inverted
STAMPING
Stamp
Pneumatic or electric output to identify the part
START PRESS
Start pressure
No flow measure taken as long as the
parameterized pressure not reached
STD CONDITIONS
Standard conditions
Calculate measured flow in other ATM pressure
and temperature condition (displays the unit with
an *)
STEP FLOW
Step flow
Minimal and maximal flow value during the step
UNITS
Units
Access to International System or American or
Custom Units
UP CONTACT
Up contact
Contact type parameter
between the inferior and
superior pressure limits during the rising, open or
close. It is the same pressure levels as for flow
event.
VALVE CODES
Valve codes
Available outputs for external automatism