ATAGO DR-M2 Standard Operating Procedure Download Page 4

 

 

 

 

P-SOP0803-1410-E01 

4/4 

Standard Operating 

Procedure  

SOP

 

 

 

 

Multi-Wavelength Abbe Refractometer 

DR-M2 

 

 

6-2. Proper care 

Depending on the operational malfunction  (abnormalities) of the DR-M2, decide on the 

proper care to correct. 

The cause of operational malfunction (abnormalities) can be divided into two areas. 

 

1)

 

Measured value becomes lower or higher due to a mistake in operation. 

a.

 

Sample from the previous measurement may have been left on the prism 

surface. 

b.

 

Distilled water was left on the prism surface. 

In case of the above malfunctions, the end user can resolve by ensuring the above 

points are checked and avoided.  Carry out operation and measurement again.  

Perform step 3 (Periodical inspection of the preceding clause). 

 

2

 

Failure in machinery or parts 

When  the  DR-M2  does not operate  properly due to mechanical trouble, under the 

same measuring conditions, degradation of the parts inside the unit and/or failure is 

assumed.  The DR-M2 may be in need of repair. 

 

a.

 

Consult with the supervisor of the refractometer. 

b.

 

Record the trouble on an operating work report. 

c.

 

Contact the trading company that you purchased the refractometer. 

d.

 

Contact  the  maker of the refractometer.  (Make a  contact list of the trading 

company where the refractometer was purchased  along with the 

manufacturer.) 

e.

 

In case of repair, decide on the method of care and put them in writing. 

 

 

7. Establishment and record of revision 

When this "standard operating procedure document" is revised, record the changes on this 

sheet and sign the approval column.   

 

 

 

Record of revision 

Approval column 

Author’s 

signature 

Approved 

by: 

     Date   

            

           Establishment 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Summary of Contents for DR-M2

Page 1: ...ece and turn the measurement knob so that the boundary line is centered at the intersection point of the cross hairs Confirm that the temperature is 20 0 0 2 The display will indicate the refractive index of the sample being measured 4 Press the SELECT key to display 1 in the mode indicator Set the function mode number to 4 with the key Press the SET key and the refractive index at the present bou...

Page 2: ...lint free tissue to remove all traces of the contact liquid Keep the instruction manual of the DR M2 for quick reference 2 Daily check Turn on the power supply switch and if 20 0 0 2 is indicated on the display of the temperature indicator perform the reference point adjustment with distilled water By performing reference point adjustment with distilled water measurements with the DR M2 can be tak...

Page 3: ...h new ones 6 In the event of a malfunction failure When operational malfunction abnormalities of the DR M2 occurs check the operation of the DR M2 by the following procedure After turning on the DR M2 check the technical malfunction by applying step 2 Daily check and adjustment with distilled water and measurement of a solution example Brix 10 nD 1 3478 of a known concentration 6 1 Check Procedure...

Page 4: ... the preceding clause 2 Failure in machinery or parts When the DR M2 does not operate properly due to mechanical trouble under the same measuring conditions degradation of the parts inside the unit and or failure is assumed The DR M2 may be in need of repair a Consult with the supervisor of the refractometer b Record the trouble on an operating work report c Contact the trading company that you pu...

Reviews: