ATAGO DR-M2 Standard Operating Procedure Download Page 3

 

 

 

 

P-SOP0803-1410-E01 

3/4 

Standard Operating 

Procedure  

SOP

 

 

 

 

Multi-Wavelength Abbe Refractometer 

DR-M2 

 

 

                      

 

4. Calibration 

   

 

1)

 

Carry out calibration once a year. 

2)

 

Perform calibration in accordance with the performance test method of Operational 

Qualification (OQ). 

3)

 

Enter the data, and other information in a performance test report, and have the 

operator performing the test and an manager confirm by signing the test results.    You 

may also request the manufacturer to perform the calibration. 

 

 

 

 

 

 

 

 

 

 

 

 

5. Maintenance 

 

1)

 

When measurement is completed, be sure that any sample on the prism surface and 

surrounding areas are completely wiped off with tissue paper moistened with  water, 

then wipe dry with a soft,lint-free tissue. 

2)

 

When the body of the instrument is dirty, wipe it off with dry tissue, cloth, etc. 

3)

 

Place the refractometer at a designated location.   If  it is not  to be used  for a long 

period of time, store the instrument in a place of relative humidity less than 75%. 

4)

 

A Desiccant case is installed within the body of the DR-M2. If the indicator inside the 

desiccant case changes from blue to pink, replace the desiccant, the indicator, and the 

cushion with new ones. 

 

 

 

6. In the event of a malfunction (failure) 

When operational malfunction (abnormalities) of the DR-M2 occurs, check the operation of 

the DR-M2 by the following procedure. 

After turning on the DR-M2, check the technical malfunction by applying step 2.  (Daily 

check) and adjustment with distilled water and measurement of a solution (example: Brix 

10%=nD 1.3478) of a known concentration.   

 

6-1. Check Procedure 

 

1)

 

Turn off the power supply switch of the DR-M2, and turn on again. 

2)

 

Perform reference point adjustment with distilled water. 

3)

 

Measure the solution (example: Brix 10%=nD 1.3478) of a known concentration.     

 

In the above procedure, operational malfunction (abnormalities) can be checked or found. 

Properly identify the situations of operational malfunction (abnormalities) of the DR-M2. 

 

 

When calibrating the Multi-Wavelength Abbe Refractometer

 

DR-M2 at your facilities (end-user), 

it is required to meet the strict conditions (maintenance, environmental conditions, examination 

conditions, the accuracy of standard liquid, and the traceability of the jigs used) for calibration. 

In order to calibrate the DR-M2 properly, enormous expense and time may be spent. 

When the  matter  of expense  is taken into consideration, an alternative is  to request the 

manufacturer to carry out calibration. 

*In this case, contact the company where the refractometer was purchased. 

 

 

 

 

 

 

     

 

 

 

  

 

 

 

    

 

Summary of Contents for DR-M2

Page 1: ...ece and turn the measurement knob so that the boundary line is centered at the intersection point of the cross hairs Confirm that the temperature is 20 0 0 2 The display will indicate the refractive index of the sample being measured 4 Press the SELECT key to display 1 in the mode indicator Set the function mode number to 4 with the key Press the SET key and the refractive index at the present bou...

Page 2: ...lint free tissue to remove all traces of the contact liquid Keep the instruction manual of the DR M2 for quick reference 2 Daily check Turn on the power supply switch and if 20 0 0 2 is indicated on the display of the temperature indicator perform the reference point adjustment with distilled water By performing reference point adjustment with distilled water measurements with the DR M2 can be tak...

Page 3: ...h new ones 6 In the event of a malfunction failure When operational malfunction abnormalities of the DR M2 occurs check the operation of the DR M2 by the following procedure After turning on the DR M2 check the technical malfunction by applying step 2 Daily check and adjustment with distilled water and measurement of a solution example Brix 10 nD 1 3478 of a known concentration 6 1 Check Procedure...

Page 4: ... the preceding clause 2 Failure in machinery or parts When the DR M2 does not operate properly due to mechanical trouble under the same measuring conditions degradation of the parts inside the unit and or failure is assumed The DR M2 may be in need of repair a Consult with the supervisor of the refractometer b Record the trouble on an operating work report c Contact the trading company that you pu...

Reviews: