ATAGO DR-M2 Standard Operating Procedure Download Page 2

 

 

 

 

P-SOP0803-1410-E01 

2/4 

Standard Operating 

Procedure  

SOP

 

 

 

 

Multi-Wavelength Abbe Refractometer 

DR-M2 

 

10

 

For measuring a  solid, the sample  piece  must be  cut and/or polished to the 

specifications. Take the sample and place onto  the prism surface of the main prism 

using contact liquid. Do not close the secondary prism onto the sample. 

11

 

Position the light guide in front of the lighting window of the main prism. The point of 

the light guide should be set to slightly higher to the main prism. 

12

 

Press the LAMP key and turn it ON. Look through the eyepiece and find the boundary 

line by gradually turning the measurement knob. Adjust the boundary line to meet the 

intersection point of the cross hairs. 

13

 

The value that is indicated on the display when the boundary  line meets the 

intersection point of the cross hairs is the measured value of the sample. 

14

 

After the operation of measurement, remove the solid sample from the prism surface 

and wipe off the contact liquid. The prism should be carefully cleaned using ethyl 

alcohol and a soft, lint-free tissue to remove all traces of the contact liquid. 

 

  *Keep the "instruction manual" of the DR-M2 for quick reference. 

 

 

 

 

2. Daily check 

 

Turn on the power supply switch, and  if 20.0(

±

0.2

) is indicated on the display of the 

temperature indicator, perform  the  reference point adjustment with distilled water.  By 

performing reference point adjustment with distilled water, measurements with the DR-M2 

can be taken consistently and accurately.    (Although it is not necessary to perform reference 

point adjustment  checks after each measurement.  It is recommended to perform reference 

point adjustment again when the refractometer has  not been used for an interval of  more 

than two hours.) 

 

3. Periodical Inspection 

 

Inspect the DR-M2 with a test piece once a month to see if it indicates the measured value 

correctly. 

Connect a circulating constant temperature bath to the DR-M2, and set the temperature to 

20.0(

±

0.2

).   

Record and save the checked data. 

 

Inspection by test piece 

The DR-M2 display indicates the refractive index. 

Measure distilled water and test piece. 

Test pieces for inspection are available for purchase. 

 

Record of Periodical inspection 

 

 

Distilled water 

(0.0%) 

Test piece 

nD 1.5164 

Test piece 

nD 1.6200 

1

st

 

measurement 

 

 

 

2

nd

 

measurement 

 

 

 

3

rd

 

measurement 

 

 

 

Summary of Contents for DR-M2

Page 1: ...ece and turn the measurement knob so that the boundary line is centered at the intersection point of the cross hairs Confirm that the temperature is 20 0 0 2 The display will indicate the refractive index of the sample being measured 4 Press the SELECT key to display 1 in the mode indicator Set the function mode number to 4 with the key Press the SET key and the refractive index at the present bou...

Page 2: ...lint free tissue to remove all traces of the contact liquid Keep the instruction manual of the DR M2 for quick reference 2 Daily check Turn on the power supply switch and if 20 0 0 2 is indicated on the display of the temperature indicator perform the reference point adjustment with distilled water By performing reference point adjustment with distilled water measurements with the DR M2 can be tak...

Page 3: ...h new ones 6 In the event of a malfunction failure When operational malfunction abnormalities of the DR M2 occurs check the operation of the DR M2 by the following procedure After turning on the DR M2 check the technical malfunction by applying step 2 Daily check and adjustment with distilled water and measurement of a solution example Brix 10 nD 1 3478 of a known concentration 6 1 Check Procedure...

Page 4: ... the preceding clause 2 Failure in machinery or parts When the DR M2 does not operate properly due to mechanical trouble under the same measuring conditions degradation of the parts inside the unit and or failure is assumed The DR M2 may be in need of repair a Consult with the supervisor of the refractometer b Record the trouble on an operating work report c Contact the trading company that you pu...

Reviews: