32
●
Inputting the temperature correction factor
①
Connect the power according to the
procedure described on page 24,
section "9.Power Supply".
②
The current Brix(%) is displayed on
the screen.
If there is no sample on the prism
surface, [LL.L] will be displayed.
③
Press the
key for one second.
[t] is displayed.
Press the
key again and the
display will change to [2] (Fig. 15-1).
④
Press the
key.
The temperature correction factor will
blink on the display (Fig. 15-2).
⑤
If "1.00" is displayed, the temperature
correction value is for sucrose.
To change this setting, press the
or
keys until the desired factor is
displayed.
⑥
Pressing the
key sets the
adjustment
of
the
temperature
correction factor and the display
returns to [2] (Fig. 15-1).
⑦
Each time the
key is pressed, the
menu item displayed switches in the
order of: [3], [4], [5], and Brix(%).
Select Brix(%).
N
If 30 seconds or more has passed
at any of the
③
through
⑥
stage, the display enters into the
continuous display mode of the
Brix(%) values.
Fig. 15-1
Example of the adjustment of the
temperature correction factor to 2.00.
Fig. 15-2