14. Adjusting to the Reference
Before adjusting to a reference solution, confirm that the prism surface is clean.
Confirm that the sample inlet unit is
properly connected to the piping.
Let tap water or a reference sample
flow through the piping.
The current Brix(%) is displayed.
display will change to [1] (Fig. 14-1).
Then, press the
and the display returns to [1] (Fig.
14-1).
menu item displayed switches in the
order of: [2], [3], [4], [5], and Brix(%).
Select Brix(%).
If 30 seconds pass when in steps
returns to the continuous Brix(%)
display.
To reset the reference point to original
factory settings
While the Brix(%) is blinking in step
above, press and hold down both the
The reference point is now reset to
original factory settings.