9. Error messages
-14-
This instrument alarms erroneous or incomplete operation by error messages.
The followings are kinds of error messages.
Standardization (Zero setting) error
●Standard liquid is not dripped on the prism surface and Standardization
(Zero setting) is made in a state of the prism surface being exposed to
the air.
●Standardization (Zero setting) is made with a sample.
Sampling error
●No sample exists on the prism surface or measurement is made in an
incomplete condition.
Range over error
●A liquid of high concentration exceeding the measuring range is
measured.
Battery error
●The battery is low. Replace the battery with new one.
●Immediately before the battery is used up, the instrument may, perform
erroneous operation, without displaying the error message "Lo". In such
a case, replace the battery with new one.
Temperature measurement range "Low Range Over Error"
●The temperature detected is lower than the measurement temperature
range (approx. 9℃ or below).
Temperature measurement range "High Range Over Error"
●The temperature detected is higher than the measurement temperature
range (approx. 51℃ or above).
External light interference
●Too much light is entering the prism, and the instrument cannot measure
accurately.
".." Standardization temperature error
●The prism is below approx.18℃ or is above approx.25℃ during
standardization.