BF5 User Manual v 2.3
Technical reference
29
Technical reference
Specifications
BF5 Accuracy and Resolution
The following accuracy figures give 95% confidence limits, i.e. 95% of
individual readings will be within the stated limits under normal climatic
conditions.
Output setting
PAR
Energy
Illuminance
Units
mol.m
-2
.s
-1
W.m
-2
klux
Overall
accuracy: Total
10
mol.m
-2
.s
-1
12%
5W.m
-2
12%
0.600 klux
12%
Overall
accuracy: Diffuse
10
mol.m
-2
.s
-1
15%
20W.m
-2
15%
0.600 klux
15%
Resolution
0.6
mol.m
-2
.s
-1
0.3 W.m
-2
0.060 klux
Range
0-2500
mol.m
-2
.s
-1
0-1250 W.m
-2
0-200 klux
Analogue output
sensitivity
1mV =
1
mol.m
-2
.s
-1
1mV =
0.5 W.m
-2
1mV =
0.100 klux
Analogue output
range
0-2500mV
0-2500mV
0-2000 mV