PC: Evolve Columns Operating Manual v2
14
Figure 5 Photograph showing the installation of the fixed bed support
3.
Locate fixed flow cell into the tube unit and fix in place.
(3) Levelling the Evolve™ Process Columns
After installation of the fixed flow cell and before installation of the adjuster unit, it is
recommended to level the column. This is achieved using the spirit level provided in the tool kit and
adjustment of the levelling studs located in the foot ring of the column.
Place the spirit level on the tube flange and adjust the levelling studs to ensure column is level
before proceeding.
There is approximately 30 mm free vertical movement available whilst still maintaining a safe
amount of thread engagement.
(4) Installation of Adjuster Bed Support
The adjuster assembly should be removed from acrylic tube unit and placed on its side on a clean
work surface.
With the rough side facing towards the ribbed flow cell, press the edge of the seal into the location
groove on the side of the flow cell. See Figure 6.
Figure 6 Diagram showing the location of the adjuster bed support