Operating notes
VLM60 User Manual
ASTECH GmbH
page 9
which is part of the delivery. It can also be downloaded from the ASTECH website. The standard baud rate of the
VLM60 is 115200baud (8N1).
For the installation of the driver, Windows must be fully started and the VLM60 must not be connected to the
PC. Now the driver installation (an executable EXE file) can be done. Afterwards the VLM60 can be connected to
the PC. It is recommended to connect the cable end to the VLM60 first and then the other cable end to the PC.
Windows should now recognize the new hardware and use the previously installed FTDI driver. When the
installation is complete, a new COM port will appear in the Ports section of the Windows Device Manager. This
COM port must be selected later when using the VLMTool for communication.
Programming cables can be obtained in various lengths from the manufacturer.
2.3
Signal-LED
There is a two-color LED in the lid of the VLM60. This indicates the current operating state of the device. Table 4
shows the possible states.
Color
Light condition
Event
Red
continuous or
blinking
Very bad measurement condition,
very bad motion detection
Orange
Continuous or
blinking
Moderate measurement condition,
moderate motion detection
Green
Continuous or
blinking
Optimal measurement condition,
optimal motion detection
Table 4: Light conditions of the signal LED
Furthermore, serious internal error faults are represented by a red pulse sequence. In this case, the manufacturer
needs to contacted.
2.4
Maintenance and Cleaning
The VLM60 is maintenance-free. For proper and long-lasting operation, the viewing window should be checked
at regular intervals and, if necessary, cleaned of dirt. If the viewing window is cracked or broken, it must be
replaced. The replacement is carried out exclusively by the manufacturer.
A damp, lint-free cloth should be used to clean the viewing window. If the viewing window is very dusty, it is
recommended to remove the dust particles with compressed air or a brush beforehand to avoid scratches on
the glass. Do not use any aggressive agents for cleaning.
If the VLMTool is installed, the user is automatically offered to install the IUSB driver as well. For this
purpose, it is recommended not to connect the VLM60 to the PC. The VLM60 may only be connected
to the PC after the VLMTool and driver have been completely installed.