TEST LEAD CONNECTIONS

If there is no ground circuit present on the DUT, you can use test leads to connect 
the OMNIA II to the DUT.

 

 Plug the black return lead (P/N 38490) into the Return terminal. Plug black 
sense lead into the Sense (-) terminal. 

 

 Attach the clip-terminated end of the return lead to the dead metal on 
the chassis of the DUT.

 

 Connect the high voltage lead (P/N 4040A-08) into the High Voltage 
terminal on the front panel of the instrument.

 

 Attach the high voltage lead to the current-carrying conductors of  
the DUT.

Did you know? 

When performing a 

Ground Bond test

 on a 

Class I product

 

connect the red high current lead (P/N 38489) into the current terminal and red 
sense lead into the sense (+) terminal. The clip at the other end will connect to the 
ground pin of the DUT plug.

Did you know? 

If necessary, a 10 x 20 cm piece of foil can be attached to the 

enclosure to simulate a full hand contact to the DUT. This foil acts as the return 
point. The return and sense (-) terminal are attached to the foil. Always check to 
make sure you made a good connection between the DUT and the return clip.

Class I Product

Class II Product

TEST

RESET

POWER

ELECTRICAL SAFETY COMPLIANCE ANALYZER

OMNIA

®

SERIES

My

Menu

1

2

ABC

DEF

GHI

JKL

MNO

PQRS

TUV

WXYZ

3

4

5

6

7

EXIT

ENTER

SCANNER STATUS

H.V.

GND

HI

LO

CH 1  2  3  4  5  6  7  8  
ON

8

9

0

< - -

CAUTION

HIGH VOLTAGE

5KV  MAX.

H.V.

SENSE+

SENSE-

CURRENT

RETURN

TEST

RESET

POWER

ELECTRICAL SAFETY COMPLIANCE ANALYZER

OMNIA

®

SERIES

My

Menu

1

2 ABC

DEF

GHI

JKL

MNO

PQRS

TUV

WXYZ

3

4

5

6

7

EXIT

ENTER

SCANNER STATUS

H.V.

GND

HI

LO

CH 1  2  3  4  5  6  7  8  
ON

8

9

0

< - -

CAUTION

HIGH VOLTAGE

5KV  MAX.

H.V.

SENSE+

SENSE-

CURRENT

RETURN

FOIL

6

7

TEST CONNECTIONS

WARNING: DO NOT TOUCH THE DEVICE UNDER TEST ONCE YOU START THE TEST.

ADAPTER BOX CONNECTIONS (8204/8254)

The adapter box (P/N 38482) provides an easy way to connect the OMNIA II to a 
device under test (DUT). 

Note:

 

If you are testing a Class I product, safety agencies may require you to perform 

a Ground Bond test on the ground circuit of the DUT along with the Hipot test.

 

 Using the output ports on the rear panel, plug the black high voltage lead in 
the High Voltage terminal.

 

 Plug the sense lead into the Sense (+) terminal. Plug the high current test 
lead into the Current terminal. These two red leads act as the ground 
connection.

 

 Connect the ground return clip (P/N 38490) to the dead metal on the chassis 
of the DUT.  Plug the sense lead into the Sense (-) terminal. Plug the return 
lead into the Return terminal.  Always check to make sure a good connection 
is made between the DUT and the return clip.

 

Plug the power cord of the DUT into the adapter box receptacle.

SENSE+

H.V.

SENSE-

CURRENT

RETURN

SIGNAL INPUT

SIGNAL OUTPUT

SCANNER 1

SCANNER 2

RS232

T10A  250V

GND

CAT. II

115V~ /230V~

10A MAX., 50Hz/60HZ

CAUTION

HIGH VOLTAGE

5KV MAX.

CAUTION

HIGH VOLTAGE

5KV MAX.

GROUND

TEST 

OUTPUT

CH

H.V.

TEST

OUTPUT

OUTPUT

REMOTE INTERFACE

INPUT

G-COM.

CH1

CH2

CH3

CH4

CH5

CH6

CH7

CH8

CAUTION

HIGH VOL

TAGE

5KV

AC MAX.

6KVDC MAX.

N

L

CAUTION

HI

GH VOL

TAGE

5KV

AC MAX.

6KVDC MAX.

ADAPTER BOX CONNECTIONS (8206/8256 & 8207/8257)

Note:

 Use the rear output terminal for connecting the adapter box (P/N 38578) on the 

8206/8256 and 8207/8257 models.

 

 Using the DUT output ports on the rear panel, plug the line lead from the 
adapter box into the L terminal and the neutral lead into the N terminal.

 

 Attach the red ground lead from the adapter box to the GND terminal.

 

 Attach the black case return (P/N CBLHC40-10TL) lead into the case terminal.

 

 Connect the ground return clip to dead metal on the chassis of the DUT.

 

Plug the line power cord from the DUT into the adapter box receptacle.



REMOTE OUTPUT

277V~MAX.  16A  MAX.

EXTERNAL

CASE

CND

PROBE HI PROBE LO

L

L

N

N

CAUTION

HIGH VOLTAGE

5KV MAX.

CAUTION

HIGH VOLTAGE

5KV

16A  AX. MAX.

DUT OUTPUTS

DUT POWER  I/P

MD CIRCUITS

SENSE+

H.V.

SENSE-

CURRENT

RETURN

SIGNAL INPUT

SIGNAL OUTPUT

REMOTE OUTPUT

277V~MAX.  16A  MAX.

EXTERNAL

CASE

CND

PROBE HI PROBE LO

L

L

N

N

SCANNER 1

SCANNER 2

RS232

T10A  250V

GND

CAT. II

115V~ /230V~

10A MAX., 50Hz/60HZ

CAUTION

HIGH VOLTAGE

5KV MAX.

CAUTION

HIGH VOLTAGE

5KV MAX.

CAUTION

HIGH VOLTAGE

5KV

16A  AX. MAX.

OUTPUT

REMOTE INTERFACE

INPUT

DUT OUTPUTS

DUT POWER  I/P

MD CIRCUITS

CAUTION

HIGH VOL

TAG

E

5KV

AC MAX.

6KVDC MAX.

N

L

CAUTION

HIG

H VOL

TAGE

5KV

AC MAX.

6KVDC

 MAX.

TEST CONNECTIONS

Summary of Contents for OMNIA II 8204/8254

Page 1: ...ch clips directly and always connect the return lead first You should always know when a test is being performed WARNING THISGUIDEWASCREATEDFOROPERATORSHAVINGSOMEFAMILIARITYWITHELECTRICALSAFETY TESTING AN ELECTRICAL SAFETY TESTER PRODUCES VOLTAGES AND CURRENTS THAT CAN CAUSE HARMFUL OR FATAL ELECTRIC SHOCK TO PREVENT ACCIDENTAL INJURY OR DEATH THESE SAFETY PROCEDURES MUST BE STRICTLY OBSERVED WHEN...

Page 2: ...C WITHSTAND Voltage HI Limit Total LO Limit Total Ramp Up Dwell Time Ramp Down Frequency Arc Detect 1240 V 10 00mA 1 0s 5 0 00mA 10 00mA 0 00mA 0 00mA 0 1s 0 0s 60Hz OFF Arc Sense HI Limit Real LO Limit Real Offset Step Name Defaults More Enter the value for each parameter by using the alpha numeric key pad Select ENTER to save and go to the next parameter 1 2 ABC 3 DEF 4 5 JKL GHI 6 MNO 7 8 TUV P...

Page 3: ...Save As Delete Load FILE SETUP Use the alpha numeric key pad to give the file a name Select ENTER to save 1 2 ABC 3 DEF 4 5 JKL GHI 6 MNO 7 8 TUV PQRS 9 EXIT ENTER WXYZ 0 My Menu CREATE FILE Caps Lock Symbol Key ON File Name My Menu File Name 0002 TEST Add Edit Delete Prompt File Fail Stop ON SETUP TESTS 0002 TEST 0001 0002 0002 0001 START From the Main Menu select the Setup Tests soft key If a te...

Page 4: ...est Using the output ports on the rear panel plug the black high voltage lead in the High Voltage terminal Plug the sense lead into the Sense terminal Plug the high current test lead into the Current terminal These two red leads act as the ground connection Connect the ground return clip P N 38490 to the dead metal on the chassis of the DUT Plug the sense lead into the Sense terminal Plug the retu...

Page 5: ... ACW Settings 1240 V 10 00mA 1 0s 0 000mA My Menu File Name 0002 TEST RESULTS 0002 LLT Settings 10000uA 0 5s 0 0uA 0001 ACW Settings 1240 V 10 00mA 1 0s 0 000mA 0002 LLT Settings 10000uA 0 5s 0 0uA PERFORM TESTS 1 2 ABC 3 DEF 4 5 JKL GHI 6 MNO 7 8 TUV PQRS 9 EXIT ENTER WXYZ 0 My Menu File Name 0002 TEST Single Step Fail Stop Load Files Results OFF ON 0001 ACW Settings 1240 V 10 00mA 1 0s 0 000mA 0...

Reviews: