INSTRUMENT SETUP
Connect the female end of the standard NEMA style power cord into the
input power receptacle on the rear panel of the instrument. Plug the male
end of the cord into a grounded power source.
Connect the interlock disable key to the signal input connector on the
rear panel of the instrument. This is required to run a test.
Turn the POWER switch to ON.
After a test has been set, you may edit the test parameters at any time.
ADD NEW TESTS
EDIT TEST PARAMETERS
WARNING: LOCATE A SUITABLE TESTING AREA WITH A CLASS I GROUNDED OUTLET.
BE SURE THAT YOUR CLASS I OUTLET HAS BEEN TESTED FOR PROPER WIRING. READ
THE SAFETY CHECKLIST OF THIS GUIDE BEFORE STARTING TO TEST.
SENSE+
H.V.
SENSE-
CURRENT
RETURN
SIGNAL INPUT
SIGNAL OUTPUT
REMOTE OUTPUT
277V~MAX. 16A MAX.
EXTERNAL
CASE
CND
PROBE HI PROBE LO
L
L
N
N
SCANNER 1
SCANNER 2
RS232
T10A 250V
GND
CAT. II
115V~ /230V~
10A MAX., 50Hz/60HZ
CAUTION
HIGH VOLTAGE
5KV MAX.
CAUTION
HIGH VOLTAGE
5KV MAX.
CAUTION
HIGH VOLTAGE
5KV
16A AX. MAX.
OUTPUT
REMOTE INTERFACE
INPUT
DUT OUTPUTS
DUT POWER I/P
MD CIRCUITS
SENSE+
H.V.
SENSE-
CURRENT
RETURN
SIGNAL INPUT
SIGNAL OUTPUT
REMOTE OUTPUT
277V~MAX. 16A MAX.
EXTERNAL
CASE
CND
PROBE HI PROBE LO
L
L
N
N
SCANNER 1
SCANNER 2
RS232
T10A 250V
GND
CAT. II
115V~ /230V~
10A MAX., 50Hz/60HZ
CAUTION
HIGH VOLTAGE
5KV MAX.
CAUTION
HIGH VOLTAGE
5KV MAX.
CAUTION
HIGH VOLTAGE
5KV
16A AX. MAX.
OUTPUT
REMOTE INTERFACE
INPUT
DUT OUTPUTS
DUT POWER I/P
MD CIRCUITS
TEST
RESET
POWER
ELECTRICAL SAFETY COMPLIANCE ANALYZER
OMNIA
®
SERIES
My
Menu
1
2
ABC
DEF
GHI
JKL
MNO
PQRS
TUV
WXYZ
3
4
5
6
7
EXIT
ENTER
SCANNER STATUS
H.V.
GND
HI
LO
CH 1 2 3 4 5 6 7 8
ON
8
9
•
0
< - -
CAUTION
HIGH VOLTAGE
5KV MAX.
H.V.
SENSE+
SENSE-
CURRENT
RETURN
MENU
Setup System
Setup Tests
Perform Tests
VERI-CKEK
Information
OMNIA II 8257
8206/8256 8207/8257
8206/8256 8207/8257
TEST SETUP
To Grounded
Power Source
2
3
START
Select the Add soft key.
From the Main Menu select
the Setup Tests soft key.
Select the a Test Type.
My
Menu
0001
Add
Edit
Delete
Prompt
File
Fail Stop ON
SETUP TESTS
My
Menu
File Name: 0001 TEST
0001
Dielectric
Withstand
Insulation
Resistance
AC Ground Bond
DC Continuity
Run Test
Line Leakage
SETUP TESTS
My
Menu
AC - WITHSTAND
Voltage
HI-Limit
Total
LO-Limit
Total
Ramp Up
Dwell Time
Ramp Down
Range :
Voltage 0V-5000V
Frequency
Arc Detect
1240 V
10.00mA
1.0s
5
0.00mA
10.00mA
0.00mA
0.00mA
0.1s
0.0s
60Hz
OFF
Arc Sense
HI-Limit
Real
LO-Limit
Real
Offset
Step Name
Defaults
More
Use the arrow keys to choose a
parameter to edit.
1
2
ABC
3
DEF
4
5
JKL
GHI
6
MNO
7
8
TUV
PQRS
9
EXIT
ENTER
WXYZ
•
0
—
<
- -
My
Menu
AC - WITHSTAND
Voltage
HI-Limit
Total
LO-Limit
Total
Ramp Up
Dwell Time
Ramp Down
Frequency
Arc Detect
1240 V
10.00mA
1.0s
5
0.00mA
10.00mA
0.00mA
0.00mA
0.1s
0.0s
60Hz
OFF
Arc Sense
HI-Limit
Real
LO-Limit
Real
Offset
Step Name
Defaults
More
Enter the value for each parameter by
using the alpha/numeric key pad. Select
ENTER to save and go to the next parameter.
1
2
ABC
3
DEF
4
5
JKL
GHI
6
MNO
7
8
TUV
PQRS
9
EXIT
ENTER
WXYZ
•
0
—
<
- -
1
2
ABC
3
DEF
4
5
JKL
GHI
6
MNO
7
8
TUV
PQRS
9
EXIT
ENTER
WXYZ
•
0
—
<
- -
My
Menu
AC - WITHSTAND
Voltage
HI-Limit
Total
LO-Limit
Total
Ramp Up
Dwell Time
Ramp Down
Frequency
Arc Detect
1240 V
10.00mA
1.0s
5
0.00mA
10.00mA
0.00mA
0.00mA
0.1s
0.0s
60Hz
OFF
Arc Sense
HI-Limit
Real
LO-Limit
Real
Offset
Step Name
Defaults
More
Select EXIT key followed by the ENTER
key to save your selection and return to
the Setup Tests screen.
1
2
ABC
3
DEF
4
5
JKL
GHI
6
MNO
7
8
TUV
PQRS
9
EXIT
ENTER
WXYZ
•
0
—
<
- -
My
Menu
MENU
Setup System
Setup Tests
Perform Tests
VERI-CKEK
Information
OMNIA II 8257
START
Use the arrow keys to choose a test
to edit within the Setup Tests screen.
My
Menu
0002 IR Settings
500V 0.00MΩ
0.5s 0.05MΩ
File Name: 0001
Add
Edit
Delete
Prompt
File
Fail Stop ON
SETUP TESTS
Use the arrow keys to choose a
parameter to edit.
1
2
ABC
3
DEF
4
5
JKL
GHI
6
MNO
7
8
TUV
PQRS
9
EXIT
ENTER
WXYZ
•
0
—
<
- -
1
2
ABC
3
DEF
4
5
JKL
GHI
6
MNO
7
8
TUV
PQRS
9
EXIT
ENTER
WXYZ
•
0
—
<
- -
Select EXIT key.
You will return to the Setup Tests screen.
0001 ACW Settings
1240V 10.00mA
1.0s 0.000mA
Select the Edit soft key.
My
Menu
0002 IR Settings
500V 0.00MΩ
0.5s 0.05MΩ
File Name: 0001
Add
Edit
Delete
Prompt
File
Fail Stop ON
SETUP TESTS
0001 ACW Settings
1240V 10.00mA
1.0s 0.000mA
My
Menu
0002 IR Settings
500V 0.00MΩ
0.5s 0.05MΩ
File Name: 0001
Add
Edit
Delete
Prompt
File
Fail Stop ON
SETUP TESTS
0001 ACW Settings
1240V 10.00mA
1.0s 0.000mA
My
Menu
INSULATION RESISTANCE
Voltage
HI-Limit
LO-Limit
Ramp Up
Dwell Time
Delay Time
Range :
Voltage 0V-6000V
500 V
0.00MΩ
0.5s
0.0s
0.05MΩ
0.000uA
0.1s
1.0s
Ramp Down
Charge-Lo
OFF
DUT Output
Step Name
Defaults
Enter the value for each parameter by
using the alpha/numeric key pad. Select
ENTER to save and go to the next parameter.
1
2
ABC
3
DEF
4
5
JKL
GHI
6
MNO
7
8
TUV
PQRS
9
EXIT
ENTER
WXYZ
•
0
—
<
- -
My
Menu
INSULATION RESISTANCE
Voltage
HI-Limit
LO-Limit
Ramp Up
Dwell Time
Delay Time
Range :
Hi-Limit 0.05MΩ - 50000MΩ 0=OFF
500 V
0.00MΩ
0.5s
0.0s
0.05MΩ
0.000uA
0.1s
1.0s
Ramp Down
Charge-Lo
Step Name
Defaults
1
2
ABC
3
DEF
4
5
JKL
GHI
6
MNO
7
8
TUV
PQRS
9
EXIT
ENTER
WXYZ
•
0
—
<
- -
My
Menu
INSULATION RESISTANCE
Voltage
HI-Limit
LO-Limit
Ramp Up
Dwell Time
Delay Time
Range :
Hi-Limit 0.05MΩ - 50000MΩ 0=OFF
500 V
0.00MΩ
0.5s
0.0s
0.05MΩ
0.000uA
0.1s
1.0s
Ramp Down
Charge-Lo
Step Name
Defaults
OFF
DUT Output
OFF
DUT Output