QUICK START
24
1. Unpack this instrument from its special shipping container. Be sure to save all packaging
materials in case you need to return it to the factory for service.
2. Locate a suitable testing area and be sure you have read all safety
instructions for the operation of the instrument and suggestions on the test
area set-up in the SAFETY section of this manual. Locate a three prong grounded outlet.
Be sure the outlet has been tested for proper wiring before connecting the instrument to it.
3. Check to be sure the correct input line voltage has been selected on the
rear panel. Either 115 volts AC or 230 volts AC. Connect the power input
plug into its socket on the rear panel of the instrument. Connect the male
end of the plug to the outlet receptacle.
Please be sure that the safety ground on the power line cord is
not defeated and that you are connecting to a grounded power
source.
4. Turn on the POWER switch located on the lower left hand side of the front panel. Upon
powering the instrument up a POWER ON SELF TEST (POST) will automatically be
performed. This test will check for the condition of the ram chips, led indicators, pcb’
s and
other critical components. All of the front panel led’
s will light temporarily then go out. In
addition, you will see the Associated Research name and Model Number briefly appear on
the LCD readout and then clear itself.
You should then see the default parameters on the LCD meter as follows:
These abbreviated parameters stand for the following:
SET:
This is the parameter settings review screen.
M1:
Indicates the memory location of the current settings.
1.0 S:
The dwell timer is set to a test duration of 1 second.
1.24KVAC:
The test voltage is set to 1,240 volts AC.
10.00 mA:
The high leakage current trip point is set to 10 milliamps.
If you wish to not use any one of these parameters you must change your parameters.
Please refer to the Operation Instructions section in this manual for further details.
WARNING
CAUTION
ARI HYPOT II
4500D VER: X.X
Set M1 1.0s
1.24KVAC 10.00mA
Current software version
displayed here