SAFETY INFORMATION
1-1
PLEASE READ
Follow all warnings to prevent possible injury or damage to the chamber. Failure to follow
all warnings may void the warranty.
Warning:
Warning:
Warning:
Warning:
Warning:
Warning:
Allow a minimum of 18" clearance in front and rear of chamber for
ventilation across air cooled condenser. The internal components
and general performance of the chamber will be impaired if it is
placed in a corner or boxed in by other equipment. Do not block the
air vent on the top or the sides of the chamber because overheating
could occur.
The power requirements are 120V 1PH 60HZ. These requirements
are noted on the model nameplate located on the back of the
chamber. If a contradiction exists, follow the power requirements on
the nameplate.
The Programmer in your chamber has been “set up” and tested at the
factory. DO NOT change any parameters. Disregard programmer
instruction manual for setup procedures. Contact factory before
changing any settings.
Do not put explosive, corrosive, or materials that may emit fumes
into the chamber
Excessive live load in chamber may impair performance of the
chamber or damage the refrigeration system.
On CO
2
or LN
2
models it is important to remove the exhaust port
plug, as these systems are high pressure. Not removing the plug
from the exhaust port will cause a dangerous condition.