background image

11

10

comfimotion.com.au

A range of ComfiMotion Bed accessories are available. In addition to the below listed accessories,  

Aidacare specialises in other complimentary products such as foam mattresses, pressure care surfaces, 

over bed tables, side lockers and adjustable day chairs. Please contact Customer Service on 1300 133 120 

for further information.

CODE 

DESCRIPTION

IMAGE

LONG SINGLE – BEA830120
KING SINGLE – BEA830130
LONG DOUBLE – BEB830132
LONG QUEEN – BEB830134
KING – BEB830136

Allure Headboard

LONG SINGLE – BEA830100
KING SINGLE – BEA830110
LONG DOUBLE – BEB830112
LONG QUEEN – BEB830114
KING – BEB830116

Chesterfield Headboard

BEA830211

Crook Handle Bed Stick

BEA830212

Bed Rail

BEA824200

Aspire ComfiMotion Transport Trolley

BEA830230

Split King Upgrade Kit

BEA830214

Simplified Handset

BEA830213

Partner Handset (used with  

Long Double and Long Queen Beds)

WARNING

Do not use accessories that have not been designed for use with ComfiMotion Beds.

5. ACCESSORIES

PLEASE READ THE CRITICAL INFORMATION BELOW:

•  Ensure the user, carer and/or equipment is positioned to avoid being trapped  

or crushed during bed adjustment.

•  Ensure the handset cord is not stretched excessively.
•  Ensure the handset and electrical leads cannot become entangled with moving 

parts of the bed.

•  DISCONNECT the power lead from the mains and stow before transporting the bed.
•  If the power lead shows any sign of damage, replace immediately.

UNDERSTANDING THE RELEVANT HANDSET

ComfiMotion Beds are powered by five-to-six electric actuators, which are controlled by the Handset. 

Adjustment occurs only while a button is pressed. Adjustment will stop when the button is released,  

or the selection reaches the end of its range.

6.1 SETTING UP YOUR HANDSET(S)

NOTE: Handset cable is attached to control box out of factory; this guide is to assist if it becomes 

knocked or removed during assembly or if a replacement is needed.
1.  Disconnect power to the bed by unplugging from wall socket
2.  Disconnect the battery backup from the bed by disconnecting the below plug from the control box

3.  Wait 10 seconds for all power to deplete from the control box
4. Disconnect the existing handset(s) from the control box (if replacing) and attach your handset(s)  

to the ports shown overleaf
NOTE: only connect handsets in the configurations shown overleaf to avoid electronic system malfunction

5.  Route the handset cable(s) carefully to the side of the bed that best suits the user/usage environment
6.  Reconnect the battery backup
7.  Reconnect power plug to wall socket
8. Wait 10 seconds
9.  Conduct the Electronic System Reset procedure (see function 13 in section 6.2 Handset Operation)  

to ensure software is calibrated to the new handset

6. SETTING UP AND OPERATING THE HANDSET

WARNING

Summary of Contents for CombiMotion ACTIV CARE BEB046740

Page 1: ...USER MANUAL ACTIV CARE...

Page 2: ...ion 14 8 Assembly Instructions 15 8 1 BEDS How to Unpack Install 8 2 HEADBOARDS How to Assemble or Mount 8 3 ACCESSORIES How to Fit to Bed 8 4 SPLIT KING UPGRADE KIT 9 Storage Movement 28 10 Cleaning...

Page 3: ...IGHT This weight represents the total maximum weight of a patient on the bed This weight 200kg should not be exceeded at any time SWL FOR BED 250kg SAFE WORKING LOAD SWL This weight represents the tot...

Page 4: ...s are equipped with all the information required to competently and responsibly operate the bed and its accessories It is recommended that the User Manual be kept in an easily accessible place for qui...

Page 5: ...ensure you have familiarised yourself with the critical information below STANDARD INCLUSIONS 4 SPECIFICATIONS INCLUSIONS ALWAYS position the bed at its lowest height when unattended to minimise the r...

Page 6: ...SCONNECT the power lead from the mains and stow before transporting the bed If the power lead shows any sign of damage replace immediately UNDERSTANDING THE RELEVANT HANDSET ComfiMotion Beds are power...

Page 7: ...ge ON OFF 9 Headrest Tilt Forwards Back 2 Legrest Up Down 4 Exit Height Raise Lower 6 Night Assist Underbed Lighting 8 WARNING Extra care should be taken whilst operating the bed functions to ensure h...

Page 8: ...to your preferred position x1 PRESET ONLY To set up your memory function first move the bed to desired positions Then hold M and Night Assist button for 5 seconds 11 Massage On Off First press turns...

Page 9: ...tion Beds are a large piece of equipment requiring special attention and careful handling at all times WARNING When opening the carton ensure it is done in a safe and careful manner DO NOT USE a box c...

Page 10: ...slots Finally place the mattress onto the bed ensuring all mattress retainers hug the sides and footend 5 Side Mattress Retainer Foot End Retainer Cut safety securing straps that hold the bed in place...

Page 11: ...TH STABILISERS For use on the ACTIV CARE BED FIXED COMPANION BED TOOLS REQUIRED Impact driver or drill 5mm attachment Phillips head screw driver 5mm Headboard Shortest side Longest end to face cushion...

Page 12: ...1 SIDE RAIL Raise bed height backrest and legrest to maximum height for ease of access 1 Bed once adjusted HI LO Raise to Max BACKREST Raise to Max LEGREST Raise to Max To adjust 5 Loosen the hand wh...

Page 13: ...bed is NOT intended to be used as a bariatric bed 8 4 SPLIT KING UPGRADE KIT ACCESSORIES SPLIT KING UPGRADE KIT BEA830230 ITEM QTY DESCRIPTION 1 4 M8 x 20 Bolt 2 2 Connecting Bracket Head 3 2 Connect...

Page 14: ...control box of the RIGHT bed Use zip ties to secure cable to bed frame if needed 9 Attach the Junction Box 4 to the RIGHT bed in position shown using the M5 bolts 7 8 4 7 JUNCTION BOX LEFT BED RIGHT...

Page 15: ...from the source by pulling on the cord DO NOT pull on the power cord or handset cord to manoeuvre the bed ALWAYS APPLY THE BRAKES when the bed is in a stationary position HANDLE WITH CARE ComfiMotion...

Page 16: ...ities all other service or repair work must be performed by qualified and trained persons or agents approved by Aidacare Overall functional check that all hand control functions are operating the bed...

Page 17: ...Aidacare on 1300 133 120 or email product aidacare com au Aidacare will not warrant the safety and or correct functioning of products where any original components have been changed or modified by non...

Page 18: ...tive qualities of the mattress Comfort preference The defect is caused as a result of the product being bent squeezed exposed to cold temperatures or damp conditions for a period of time causing the m...

Page 19: ...AIDACARE PTY LTD Building 3A 1 Moorebank Avenue Moorebank NSW Australia 2170 1300 133 120 aspirecare com au MADE IN PRC ASUM80 28 250822...

Reviews: