SECTION 8 SERVICING THE BIS VISTA MONITORING SYSTEM
8-13
8.
Set the screws in the holes. Slowly tighten to 45 in-oz.
8.9
Calibrating the Touch Screen
The Touch Screen is calibrated at the factory and does not need to be re-calibrated on a
regular basis. However, if necessary, the touch screen may be calibrated by selecting
[Calibrate Touch Screen]
from the Maintenance Menu. Follow the on-screen instructions.
8.10
Using the Reset Button
The Reset button is located on the back panel of the monitor. If necessary, the software can
be reset by accessing this button with a ballpoint pen, paper clip or other similar tool.
8.11
BISx Checkout and Safety Tests
1.
Perform DSC Self Test. See Section 7.2.2 “The BISx Checkout Procedure” or refer
to the BIS VISTA Monitoring System Operating Manual.
2.
Verify that all test sections PASS.
3.
Install a known good PIC cable.
4.
Connect a Sensor Simulator or Test Sensor tool as described in Section 10
“Appendix I.”
5.
Perform a Patient Interface Cable (PIC) Checkout Procedure as described in Section
7.2.3. “PIC Checkout Procedure.” For more detail, refer to the BIS VISTA
Monitoring System Operating Manual.
6.
Perform a leakage (electrical safety test) according to the appropriate institution
requirements.
7.
If appropriate, perform a Hipot test according to institution requirements.
8.
End of Procedure.